Unpacking:
- Please unpack the device carefully
- In the case of any damage a detailed report must be sent
immediately (post, rail or forwarder).
Transport safety
Remove the transport protection un-
der the
buoyage
(4).
Correct use
Use:
Use ICC (Immersion Circulator Compact) for heating and cir-
culating fluids.
Intended Use: Tabletop device
Range of use (indoor use only):
- Laboratories - Schools
- Pharmacies - Universities
This device is suitable for use in all areas except:
- Residential areas.
- Areas that are connected directly to a low-voltage supply
network that also supplies residential areas.
The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not sup-
plied or recommended by the IKA
®
.
- If the device is operated improperly or in contrary to the
IKA
®
specifications.
- If the device or the printed circuit board are modified by
third parties.
Unpacking
Delivery scope:
- ICC baisc
- Mains cables
- Screwdriver (use for safety circuit)
- Operating instructions
- Warranty card.
Preparations
Setting up:
- Securely fix the immersion circulator on the bath.
- Place the device on an even, stable, clean, nonslip, dry
and fireproof surface.
- Keep at least 20 cm of open space at the front and rear
side.
- When a plastic bath is used, please ensure that the heater
does not contact the bath.
- The place for installation should be large enough and pro-
vide sucient air ventilation to ensure the room does not
warm up excessively because of the heat from device
radiates to the environment.
- Do not set up the device in the immediate vicinity of heat
sources and do not expose to sun light.
Low Level
High Level
Filling and draining:
- Connect the mains plug and turn on the device with
mains switch (2).
- The low level warning message appear on the display.
- Fill fluid to the bath.
Note: Pay attention to the fluid level information on the
display:
For draining, please refer to the bath operating instructions.
NOTICE