Any malfunctions during operation will be identified by an error message on the display (only RV 10 digital).
Once a serious error message has been displayed, the lift moves to the top end position and the device can no longer be operated. The lift can be
operated again.
Proceed as follows in such cases:
)Switch off device using the main switch
)Carry out corrective measures
)Restart device
Error Effect Cause Correction
code
E01 No or too little rotation Cable break Reduce the volume in the evaporator piston
Load too large
Motor blocked
E02 No communication with the Interface hidden or dirty Check / clean the interface
heating bath
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA
®
service department,
- send the device for repair, including a short description of the fault.
RV 10.2029 NS 24/40 Distillation spider with 6 sleeves
RV 10.2030 NS 24/40 Distillation spider with 12 sleeves
RV 10.2031 NS 24/40 Distillation spider with 20 sleeves
RV 10.610 Distilling sleeve 20 ml
RV 10.606 NS 29/32 Distillation spider with 5 flasks 50 ml
RV 10.607 NS 29/32 Distillation spider with 5 flasks 100 ml
RV 10.608 NS 24/29 Distillation spider with 5 flasks 50 ml
RV 10.609 NS 24/29 Distillation spider with 5 flasks 100 ml
RV 10.2032 NS 29/42 Distillation spider with 5 flasks 50 ml
RV 10.2033 NS 29/42 Distillation spider with 5 flasks 100 ml
RV 10.2034 NS 24/40 Distillation spider with 5 flasks 50 ml
RV 10.2035 NS 24/40 Distillation spider with 5 flasks 100 ml
RV 10.3000 Extension plate
RV 10.8001 Seal, PTFE compound
PC 1.1 Cable
labworldsoft
®
Error codes

Table of Contents

Related product manuals