40  tina83e1-d (2018-12) 
For adjusting the zero, operate the gauge under the same con-
stant ambient conditions and in the same mounting orientation 
as normally. 
The zero can be adjusted via 
•  the <ADJ> button on the gauge, 
•  the diagnostic port (→  [1]), 
•  the RS232C/485C interface (→  [1]), 
•  the EtherCAT interface (→  [2]). 
 
  If you are using a seal with centering ring and filter, check 
that they are clean or replace them if necessary 
(→ "Deinstallation"). 
  Put the gauge into operation and operate it at atmospheric 
pressure for at least 10 minutes. 
  Briefly press the <ADJ> button with a pin (max. ø1.1 mm) 
and the ATM adjustment is carried out, or … 
ATM
 
… perform the adjustment via the diagnostic port or via the 
EtherCAT interface. 
 
The Pirani sensor is adjusted to 1000 mbar (duration ≈5 s).