tina56e1-f (2015-09) 1
Pirani Capacitance Diaphragm
Gauge
PCG550
PCG552
PCG554
Operating Manual
Incl. EU Declaration of Conformity

Other manuals for Inficon PCG550

Questions and Answers

  • L
    Leslie GuerreroAug 3, 2025
    What to do if FAIL SENSOR lit solid red on Inficon Measuring Instruments?
    • J
      Jason JensenAug 3, 2025
      If the FAIL SENSOR light is solid red on your Inficon Measuring Instruments, it indicates that the electronics unit may not be compatible with the sensor. In this case, you should replace the sensor or replace the gauge.
  • A
    Alexandra WaltersAug 9, 2025
    What to do if FAIL EEPROM is blinking red on Inficon Measuring Instruments?
    • M
      Michael MendezAug 9, 2025
      If the FAIL EEPROM is blinking red on your Inficon Measuring Instruments, try turning the gauge off and on again after 5 seconds to reset it. If the problem continues, replace the gauge.
  • G
    Gregory HunterAug 15, 2025
    What to do if FAIL PIR1 lit solid red on Inficon Measuring Instruments?
    • J
      Jay AllenAug 15, 2025
      If the FAIL PIR1 light is solid red on your Inficon Measuring Instruments, it could be due to a defective Pirani sensor or incorrect mounting of the electronics unit on the sensor. Check the connections between the electronics and the sensor. If the connections are correct, consider replacing the sensor.
  • L
    Lori HerringAug 21, 2025
    What to do if FAIL CAP1 lit solid red on Inficon Measuring Instruments?
    • K
      Kayla PetersonAug 21, 2025
      If the FAIL CAP1 light is solid red on your Inficon Measuring Instruments, this may be caused by a defective diaphragm sensor or the electronics unit not being correctly mounted on the sensor. Ensure that the connections between the electronics and the sensor are correctly installed. If not, consider replacing the sensor.
  • C
    Courtney YoungAug 27, 2025
    What to do if FAIL ATM1 lit solid red on Inficon Measuring Instruments?
    • G
      Glenn HullAug 27, 2025
      If the FAIL ATM1 light is solid red on your Inficon Measuring Instruments, it indicates that the atmospheric pressure sensor is defective. The suggested solution is to replace the gauge.

Summary

Product Identification and Validity

Intended Use and Functional Principle

Safety Information

Safety Symbols, Personnel Qualifications, and General Instructions

Explains safety symbols, warnings, personnel skills, and general safety guidelines.

Technical Data

Installation Guide

Vacuum Connection Procedures and Warnings

Instructions for connecting the gauge to a vacuum system, including warnings.

Power Connection Procedures and Safety

Instructions for connecting the device to power, including safety precautions.

Operation

Switching Functions and Setpoint Configuration

Setting and configuring switching functions SP1 and SP2, and adjustment procedures.

Deinstallation

Explicit Device Address Setting and Deinstallation Procedure

Setting EtherCAT address and initial deinstallation steps.

Deinstallation Procedure and Warnings

Final steps for removing the gauge and associated warnings.

Maintenance and Repair

Gauge Adjustment and Calibration

Initial steps and general procedures for adjusting the gauge.

Troubleshooting Gauge and Sensor Errors

Diagnosing and correcting common gauge and sensor errors.

Sensor Replacement

Step-by-step guide for replacing the sensor unit.