tina60e1-a (2011-04) 1
Pirani Standard Gauge
PSG550
PSG552
PSG554
Operating Manual
Incl. EC Declaration of Conformity

Questions and Answers

  • S
    sspencerAug 3, 2025
    What to do if the Inficon Measuring Instruments output signal is permanently 0V?
    • M
      Mrs. Morgan MurrayAug 3, 2025
      If the output signal is permanently 0V, there might be several reasons. First, check the sensor cable for defects or incorrect connections. Second, ensure that the power supply is turned on. Third, if there is an error, remedy it. Finally, the gauge might be in an undefined state; in this case, turn the gauge off and on again after 5 seconds to reset it.
  • M
    mrasmussenAug 10, 2025
    What to do if Inficon PSG550 Measuring Instruments shows FAIL EEPROM?
    • E
      Erin LloydAug 10, 2025
      If the device displays 'FAIL EEPROM', it indicates an EEPROM error. You can try turning the gauge off and on again after 5 seconds to reset it. If the issue persists, consider replacing the gauge.
  • M
    Melissa RiosAug 19, 2025
    Why Inficon PSG550 shows FAIL SENSOR?
    • M
      Melissa PonceSep 12, 2025
      If the device displays 'FAIL SENSOR', it could be due to the electronics unit not being compatible with the sensor, in which case you should replace the sensor. If that's not the issue, consider replacing the entire gauge.
  • T
    Thomas SpenceAug 27, 2025
    What to do if Inficon PSG550 Measuring Instruments shows FAIL PIR1?
    • D
      Daniel ArnoldAug 27, 2025
      If the device displays 'FAIL PIR1', it could be due to a defective Pirani sensor, consider replacing the sensor. Also, check the connections between the electronics unit and the sensor to ensure they are correctly mounted.

Summary

Product Identification

Validity

Intended Use

Trademark

Patents

Scope of Delivery

Maintenance, Repair

Adjusting the Gauge

Calibrate the gauge for accurate pressure readings under constant conditions.

Replacing the Sensor

Guide for replacing the sensor if it's contaminated or malfunctioning.

Troubleshooting

Diagnose and resolve common problems encountered with the gauge.

Returning the Product

Disposal

Accessories

Spare Parts (Tungsten Filament)

Spare Parts (Nickel Filament)

Spare Parts (Al2 O3 Coated Filament)

Further Information

ETL Certification

EC Declaration of Conformity

Standards

List of harmonized and international standards the gauge complies with.

Manufacturer; Signatures

Details of the manufacturer and signatory personnel.

Inficon PSG550 Specifications

General IconGeneral
TechnologyPirani
Accuracy±15% of reading
Number of Sensors1
Output Signal0…10 V DC, linear
Materials exposed to vacuumStainless steel