34
tina60e1-a (2011-04)
4 Operation
When the supply voltage is applied, the measurement signal is
available at the connector ( "Power Connection").
Allow a stabilization period of at least 10 minutes. It is advisable
to operate the gauge continuously, irrespective of the pressure.
The gauge is factory calibrated. Due to long time operation or
contamination, a zero drift could occur. Periodically check the
zero and adjust it if necessary (adjusting the gauge 51).
4.1 Status Indication and Displays
Light-emitting diodes (LEDs)
SP1SP2
ADJ
ST
DIA
Status
LED State Meaning
<ST> off no supply voltage
lit green measurement mode
lit red error
<SP1> lit green Relay SP 1 closed
off Relay SP 1 open
<SP2> lit green Relay SP 2 closed
off Relay SP 2 open