37
Technical Specification
36
Test Principle and Performance
Sample type
Sample volume
Test Range
Reading time
Calibration
Operating Temperature
Operating Humidity
Strip storage temperature
Display Type
Dimension
Weight
Power source
Battery Life
Memory Capacity
PC communications port
Capillary whole blood
1.5
㎕
10-600 mg/dL (0.6-33.3mmol/L)
9 seconds
Plasma - Equivalent
10-40 °C (50-104
°F)
10-90%
2-30 °C (35.6-86
°F)
LCD
77 x 46 x 19 mm
42g (including battery)
3 V Li Battery (CR2032) x 2
Running 5,000 tests
200 test results
Communication is available to
connect computer with cable
Measurement Range :
The measurement range of the finetest Auto-coding
TM
systems is 10 to 600
mg/dL (0.6 - 33.3 mmol/L).
Test Principle :
Glucose in the blood sample will react to the electrodes in the test strip,
generating an electrical current that will stimulate a chemical reaction. This
reaction is measured by the finetest Auto-coding
TM
Meter and displayed as your
blood glucose result.
Note: Different levels of reactions will occur depending on the amount of
glucose in the blood sample.
Calibration :
The finetest Auto-coding
TM
system is plasma-calibrated to allow easy
comparison of results with laboratory methods.
Accuracy :
Studies have shown that the test results of finetest Auto-coding
TM
systems
compare well with laboratory methods.
Precision :
The precision of finetest Auto-coding
TM
System has been assessed by studies in
accordance with ISO 15197.
Results for venous blood samples and Control solutions :
Blood 43 mg/dL 2,39 mmol/L SD = 1.4 CV = 3.4%
Blood 91 mg/dL 5,05 mmol/L SD = 2.2 CV = 2.4%
Blood 145 mg/dL 8,05 mmol/L SD = 4.2 CV = 2.9%
Blood 223 mg/dL 12,38 mmol/L SD = 7.7 CV = 3.4%
Blood 360 mg/dL 19,98 mmol/L SD = 6.4 CV = 1.8%
Within Run Precision
Control 41 mg/dL 2,28 mmol/L SD = 1.5 CV = 3.8 %
Control 113 mg/dL 6,27 mmol/L SD = 3.0 CV = 2.7 %
Control 328 mg/dL 18,20 mmol/L SD = 9.3 CV = 2.9 %
Total Precision