Before You Test
Turn on Beeping sound
User’s testing option
User’s ID
The Blood Glucose Monitoring System has a wide variety of
functions to choose from. There is a beeping sound to alert you when a function
is set; date, time, five daily alarms, unique user ID, ability to designate activities
related to the test result, and the ability to designate three unique data averages.
•Press power button
and hold for
about 3 seconds
•After pressing and
releasing either or
buttons to turn the
beeper on/off, enter
•After pressing and
releasing either or
buttons to turn the
user testing option
on/off, enter
•After pressing and
releasing either or
buttons to turn user ID
on/off, enter
Date/Time Test unit
•
After pressing and releasing
either or buttons to set
the date and time, enter
•
After pressing and
releasing either or
buttons to set the test unit
of result, enter
Blood Glucose
Test Strip Information
12
Limitations of Blood glucose measuring system :
The blood glucose test Strips provide accurate results when the following constraints are
observed :
- Use the blood glucose test strips with the blood glucose
test meter or blood glucose test meter.
- Use fresh capillary whole blood only. Do not use plasma or serum
- Do not use neonate samples.
- The blood glucose test strips are for single use only. Do not reuse.
- Dehydration may lower test results. If you are severely dehydrated, contact
you physician immediately
- Inaccurate results may occur when in shock, hypotensive, hyperglycemic,
or hyperosmolar state, with or without ketosis.
- Blood glucose test strips used above altitudes of 9,840 feet will have an
effect on test results.
Doctors - Please note the following interferences that may affect test results :
- test strip results can be used with hematocrit levels in the range of 20
%
to 60%. If the hematocrit range is out of the range(20%~60%), then the test results
may be lower or higher than actual value. If you do not know your hematocrit level,
consult your healthcare professionals.
- Interferences : Acetaminophen, Uric and, Ascorbic acid (Vitamin C), and other
reducing substances when occurring in normal blood or normal therapeutic
concentrations do not significantly affect results. However, abnormally high
concentrations in may cause inaccurately high results.
- Lipemic samples; Cholesterol up to 500 mg/dL or triglyceride up to 2890 mg/dl
do not significantly affect the results. Values beyond these levels should be
interpreted with caution.
- Blood samples that contain a high concentration of dissolved oxygen may
lower the test result.
Performance Characteristics :
The performance of the blood glucose test strips has been evaluated in
laboratory and in clinical tests.
Measurement Range : The measurement range of the Blood
glucose measuring system is 10 to 600 mg/dL (0.6~33.3mmol/L).
13