Storage of your GluNEO
®
Lite System
To prevent the meter and test strips from getting dirt, dust or other contaminants, please
wash and dry your hands thoroughly before use
Meter Storage
Storage condition: 2~30 or 36~86 (Temperature)
Always store and transport the meter in its original storage case.
Avoid dropping and strong impact.
Avoid direct sunlight and humidity.
Test Strip Storage
Storage condition: 2~30 or 36~86 (Temperature)
Store your test strips in their original vial only. Do not store the test strips in other
containers.
Store test strip packages in a cool and dry place. Keep away from direct sunlight and heat.
Do not freeze.
After removing a test strip from the vial, immediately close it tightly.
Record the discard date (date vial opened, plus 6 months) on the test strip label.
Do not use if it is expired(2 years) after manufacturing.
Touch the test strip with clean and dry hands.
Do not use if 6 months have been passed, since the opening of the vial cap for test strips.
Control Solution Storage
Storage condition: Store in a cool, dry place at between 8~30 or 46~86
(Temperature)
After use, close cap tightly.
Make sure the control solution bottle is tightly closed
Record the discard date (date vial opened, plus 3 months) on the control solution on the
bottle.
Do not use if expiry date has passed.
It expires in 24 months after manufacturing date.
It expires in 3 months after opening the vial cap.
CAUTION
1. Store the GluNEO
®
Lite test strip vials in a cool, dry place. Keep out of direct
sunlight. Do not freeze.
2. Store test strips in their original vial only. Do not mix the test strips in new vials or in
any other container.
3. Immediately replace the vial cap and close tightly after removing any test strips
from the vial.
4. Make a note of the discard date, which is six months from the date you first open a
new vial of test strips. Throw GluNEO
®
Lite test strips and vial away after the
discard date.
5. Do not use the test strips after the expiration date printed on the package or vial,
Testing with expired strips it may cause inaccurate results.
6. Do not test at temperatures below 10(50°F) or above 40(104°F).
7. Do not test with humidity below 10% or above 90%.
8. Do not bend, cut, or alter the test strip.
9. Avoid getting dirt, food, and water on the test strip with wet hands.
10. Avoid getting dirt, food, and water on the color-coding label (backside of test strip)