After each decontamination process, wipe down
the case of the ProXcide™ main unit with a dry
cloth to remove any residual condensation. Only
use mild cleaning products such as normal
detergent, or mild alcohol wipes to clean the
plastic parts of the system.
Once a week, check the mains lead and the
process monitor leads for damage, paying
particular attention to the points where they
emerge from the connectors. The ProXcide™
system must be PAT (Portable Appliance
Testing) tested for electrical safety every six
months. The wheels of the ProXcide™ system
should be periodically cleaned.
The main unit of the ProXcide™ has perforated
panels / grilles that require regular cleaning with
a damp cloth to remove dust accumulation..
Do not tip the unit to achieve this, as this may
result in ProXcin™ spillage.
Section 6: ProXcide
™
maintenance
The ProXcide™ should be serviced every 6
months by Inivos. The system will flash up a
warning message for 30 seconds at power up
when this service is due.
Failure to have the unit serviced every 6
months may lead to poor decontamination
and frequent process failures.
It may be necessary to drain the fluid stored in the
ProXcide™. For example, when the vapourising unit
is to be disposed of at end of life, or to transport the
ProXcide™ by air freight.
Fluid drainage may only be performed by a
Inivos Engineer as it requires a pump and the
removal of equipment covers. It is not safe for
an Operator to perform this operation.
6.1 Regular care
6.2 ProXcide™ servicing
6.3 Draining the ProXcide™