INSTRUCTION MANUAL REVISION 4.0 APRIL 2020
Once the ProXcide
is plugged in and turned on at the wall it will run an initiation program. Coloured
LED lights provide a visual indication as to the status of the ProXcide
as it runs through a series of
checks to ensure that the process is safe to proceed. If all checks are successful, the main lights will
change from white to blue to indicate the ProXcide software has loaded and has passed all checks.
The fluid level is indicated by blue lights. The fluid refill indicator will glow red if a replacement
cartridge is required. (ProXcin cartridge change instructions are provided in Section 5.1.10.)
Pressing the single button on the machine at this stage will result in the lights changing from blue to
green to indicate the ProXcide is ready for subsequent activation from the process monitor.
In the unlikely event of interference such as power outage, the system will ‘fail safe.
The decontamination cycle is activated from the process monitor. This cycle, which is controlled
and monitored computationally onboard the
ProXcide, will then follow the steps illustrated
below:
4
SURFACE TREATMENT
The ProXcide™ continually
monitors and adjusts the
process to ensure suicient
and eicient treatment of
all surfaces..
1
PRE-START CHECKS
ProXcide™ equipment
monitors humidy, checks
fans, fluid level and
injection module.
2
START CYCLE
PREPARATION
The ProXcide™ automatically
prepares the system to start
the decontamination
process.
6
SYSTEM RESET
The ProXcide™ resets
in readiness for the
next process.
3
VAPOUR DIFFUSION
The ProXcide™ generates
and circulates the
vapour until the required
concentration has
been achieved.
5
DEACTIVATION
The ProXcide™ quickly
reduces the concentration
of atmospheric hydrogen
peroxide until deactivation
is complete.
A one button
automatic
process with
no operator
programming
3.3 Pre-start automated checks