Instruction Manual CVM211 Stinger
InstruTech Page 20
7.1 Non-Linear Analog Output
The following table is for use with CVM211 Stinger part numbers ending with the letter “NL” providing a non-
linear analog output.
Non-Linear analog output for CVM211GAA-B-NL through CVM211GHA-B-NL
True
Pressure
N
2
Ar
He
O
2
CO
2
KR
Freon12
Freon22
D
2
Ne
CH
4
0 x 10
-3
mbar
0.3751
0.375
0.375
0.375
0.375
0.375
0.375
0.375
0.375
0.375
0.375
0.1 x 10
-3
mbar
0.3759
0.3757
0.376
0.376
0.376
0.376
0.376
0.376
0.376
0.3757
0.3766
0.3 x 10
-3
mbar
0.3768
0.376
0.377
0.377
0.377
0.377
0.378
0.378
0.377
0.3763
0.378
0.6 x 10
-3
mbar
0.3795
0.378
0.379
0.38
0.381
0.377
0.382
0.381
0.381
0.3782
0.3825
1.33 x 10
-3
mbar
0.384
0.381
0.382
0.384
0.385
0.379
0.388
0.388
0.386
0.381
0.3896
2.67 x 10
-3
mbar
0.3927
0.387
0.389
0.392
0.395
0.384
0.401
0.4
0.396
0.388
0.403
6.67 x 10
-3
mbar
0.4174
0.403
0.409
0.417
0.412
0.395
0.437
0.432
0.425
0.405
0.438
13.3 x 10
-3
mbar
0.4555
0.429
0.441
0.453
0.462
0.415
0.488
0.48
0.47
0.433
0.492
26.6 x 10
-3
mbar
0.5226
0.477
0.497
0.521
0.536
0.451
0.581
0.566
0.549
0.484
0.584
66.6 x 10
-3
mbar
0.6819
0.595
0.637
0.679
0.705
0.544
0.778
0.764
0.727
0.608
0.796
0.13 mbar
0.878
0.745
0.814
0.868
0.9
0.668
1.009
0.99
0.944
0.768
1.053
0.26 mbar
1.1552
0.962
1.068
1.141
1.179
0.847
1.315
1.291
1.265
1.002
1.392
0.66 mbar
1.6833
1.386
1.589
1.664
1.668
1.194
1.826
1.805
1.914
1.469
2.014
1.33 mbar
2.2168
1.818
2.164
2.195
2.172
1.536
2.257
2.247
2.603
1.976
2.632
2.66 mbar
2.8418
2.333
2.939
2.814
2.695
1.921
2.647
2.666
3.508
2.631
3.313
6.66 mbar
3.6753
3.028
4.387
3.672
3.316
2.429
3.029
3.09
5.059
3.715
13.3 mbar
4.2056
3.48
5.774
4.225
3.67
2.734
3.204
3.33
6.361
4.605
4.699
26.6 mbar
4.5766
3.801
7.314
4.62
3.903
2.966
3.308
3.414
5.406
5.172
66.6 mbar
4.8464
4.037
4.916
4.071
3.075
3.43
3.509
6.159
5.583
133 mbar
4.9449
4.122
5.026
4.154
3.134
3.618
3.66
6.483
5.72
266 mbar
5.019
4.192
5.106
4.336
3.269
3.827
3.883
6.661
5.86
400 mbar
5.1111
4.283
5.2
4.502
3.384
3.938
4.005
6.726
533 mbar
5.2236
4.386
5.315
4.621
3.466
4.016
4.088
6.767
6.103
666 mbar
5.3294
4.477
5.422
4.708
3.526
4.076
4.151
6.803
800 mbar
5.4194
4.55
5.515
4.775
3.573
4.124
4.203
6.843
6.342
933 mbar
5.4949
4.611
5.592
4.83
3.613
4.166
4.247
6.89
1.01 Bar
5.534
4.643
5.633
4.86
3.632
4.19
4.271
6.92
1.06 Bar
5.5581
4.663
5.658
4.877
3.645
4.203
4.286
6.942
6.519
1.19 Bar
5.6141
4.706
5.713
4.919
3.674
4.237
4.321
7
1.33 Bar
5.6593
4.745
5.762
4.955
3.69
4.270
4.354
7.056
6.642
Values listed under each gas type are in volts.
Note: By design, these values are identical to the outputs from MKS Instruments / Granville-Phillips®
Convectron® gauges, Mini-Convectron® modules and controllers so that equivalent units can be interchanged
without affecting your process system or software.
An analog output of less than 0.01 volts to near 0 volt indicates a damaged or faulty sensor.
Refer to the next page if you have a CVM211 Stinger with a log-linear analog output.