Please read and comply with all precautions listed in the General Warnings section of this manual.
The Xenon Lamp Module should be replaced aer 650 hours of use or if the lamp fails to start when the power is turned on.
Xenon Lamp Module Replacement
Note
Please adhere to appropriate safety precautions when performing lamp replacement. Only qualified Personnel should service this
device. Protective face mask and gloves should be worn when replacing lamp module. Before changing the lamp module, turn power
off and allow the Light Source to cool for at least fieen (15) minutes.
Note
You must use ONLY an Integra supplied lamp module assembly for replacement in this Light Source. Use of other lamps or lamp
assemblies will void all warranties for this product. Refer to Parts List later in this manual.
To remove the lamp:
1. Make sure power switch is turned off. Disconnect the power
cord from the hospital grade receptacle.
2. Allow the unit to cool down for a minimum of 15 minutes
before replacing the lamp module. The lamp module may
represent a burn hazard if not allowed to cool sufficiently
prior to servicing the device.
3. Remove the two side screws and loosen the two quarter turn
screws in the back of the unit. See Figure 9. Pull back on the
top cover handle as indicated in Figure 10A and then pull up
on the top cover to open as indicated in Figure 10B.
4. Firmly grasp the lamp module by the protective cover with
one hand while pressing the release buon with the other.
Pull up to remove the assembly. See Figure 11. Do not use
excessive force or prying tools.
9
10a
10b
Rear panel fasteners