Integra
®
Luxtec
®
9300XSP 300 Wa Xenon Light Source
14
Electromagnetic Compatibility (EMC) User Information
Warning
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service
according to the Electromagnetic Compatibility [EMC] information provided in the accompanying documents.
Warning
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
Warning
The equipment or system should not be used adjacent to or stacked with other equipment; if adjacent or stacked use is
necessary, the equipment or system should be observed to verify normal operation in the configuration in which it will
be used.
Note
The EMC tables and other guidelines that are included in the Instruction Manual provide information to the customer
or user that is essential in determining the suitability of the Equipment or System for the Electromagnetic Environment
of use, and in managing the Electromagnetic Environment of use to permit the Equipment or System to perform its
intended use without disturbing other Equipment and Systems or non-medical electrical equipment.
Table 201: Guidance and Manufacturer’s Declaration – Emissions All Equipment and Systems
Emissions Test Compliance Electromagnetic Enforcement – guidance
RF Emissions CISPR 11 Group 1 The XSP uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF Emissions CISPR 11 Class B
Radiated and Conducted Emissions
The XSP is suitable for use in all establishments
including domestic, and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Conducted Emissions Tests Performed at both 240VAC,
50Hz and 120VAC, 60Hz
Harmonics
IEC 61000-3-2
n/a Equipment intended for Professional Use Only
Flicker
IEC 61000-3-3
n/a Equipment intended for Professional Use Only
The XSP light source is intended for use in the electromagnetic environment specified below. The customer or user of the XSP
light source should assure that it is used in such an environment.