EasyManua.ls Logo

Integra Luxtec 9300XSP - Troubleshooting; Maintenance and Cleaning; Maintenance; Cleaning

Integra Luxtec 9300XSP
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Integra
®
Luxtec
®
9300XSP 300 Wa Xenon Light Source
12
Troubleshooting
Trouble Area Possible Cause Corrective Action
No power Light source not plugged in Plug in light source
Top cover not closed Close and secure top cover
No light output Light source not turned on Turn power on
Bad/no lamp Check lamp seating/replace lamp module
Aenuator closed Check position of knob on front panel
Turret mispositioned Rotate turret to desired adapter fiing
Blown fuse Replace fuse as indicated in maintenance section
Reduced light output Cable mismatched to turret Rotate turret to matching adapter fiing
Aenuator mispositioned Check position of knob on front panel
Bad lamp Replace lamp module
Maintenance:
If the Light Source does not operate properly when connected to a grounded receptacle, check the fuses. Do not aempt to
repair the unit if the lamp fails in use. Turn off the unit and allow it to cool for at least 15 minutes, then try to restart the unit. If
the lamp still fails to illuminate, the lamp may be defective, the lamp has exceeded its useful life (see lamp replacement), or there
is a power failure. Try replacing the lamp module. If you still experience difficulties, return the unit to Integra for evaluation.
Cleaning:
The Light Source can be cleaned and disinfected using 70% isopropyl alcohol. Unplug the power cord before cleaning.
Allow 5 minutes for alcohol to evaporate before reconnecting to power.
To replace the fuse:
The fuses for the Light Source are located in the power entry module in the rear of the unit. Plug in power cord.
Using a small flat screwdriver, pry out the red plastic back from the power entry module. Check to see if either fuse is blown,
then replace with fuse(s) of the same rating (T6.3A 250V). To replace the red block in the housing, snap the retaining ring in place.
Plug cord back into Light Source and retest the unit. If unit now works, replace the cord lock with screws previously removed to
secure cord to unit.
Maintenance And Cleaning