16
Table 204: Guidance and Manufacturer’s Declaration – Emissions Equipment and Systems
that are NOT Life-Supporting
Table 206: Recommended Separation Distances Between Portable and Mobile RF Communi-
cations Equipment and the 9300XSP Equipment and Systems that are NOT Life-Supporting
Electromagnetic Compatibility (EMC) User Information
The XSP light source is intended for use in the electromagnetic environment specified below. The customer or user of the XSP
light source should assure that it is used in such an environment.
The XSP is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer
or user of the XSP can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF Communications Equipment and the XSP as recommended below, according to the maximum output power of the
communications equipment.
Immunity Test
IEC 60601 Test Level Compliance Level
Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
3 Vrms from
150 kHz to 80 MHz
V1 = 3 Vrms Portable and mobile RF communications equipment
should be separated from the 9300XSP light source
by no less than the recommended separation
distances calculated/listed below:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
E1 = 3V/m
Where P is the maximum power rating in was and D
is the recommended separation distance in meters.
Field strengths from fixed transmiers, as determined
by an electromagnetic site survey, should be less that
the compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment
containing a transmier
Recommended Separation Distances for the 9300XSP (meters)
Maximum Output Power
(Was)
150 kHz to 80 MHz 80 to 800MHz
800 MHz to 2.5 GHz
0.01 0.11667 0.11667 0.23333
0.1 0.36894 0.36894 0.73785
1 1.1667 1.1667 2.3333
10 3.6894 3.6894 7.3785
100 11.667 11.667 23.333
80 to 800 MHz
800 MHz to 2.5 GHz