D-0121151-B – 2018/11
AD528 - Instructions for Use - EN Page 42
To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is essential to use
only the following accessories:
Item
Manufacturer
Model
Monitor Headset w. microphone
Sennheiser
PC131/PC3
Bone Conductor
Radioear
B71W
Audiometric Headset
Radioear
DD65
Talk Back Microphone Clip-On
G-Star
TC-945
Free field speakers
Edifier
MP210
Patient response switch
Radioear
APS3
USB cable
Sanibel
8011241
Conformance to the EMC requirements as specified in IEC 60601-1-2 is ensured if the cable types
and cable lengths are as specified below:
Description
Length (m)
Screened (Yes/No)
Monitor Headset
w. microphone
2.9
Yes
Bone Conductor
2.0
No
Audiometric Headset
2.0
Yes
Talk Back Microphone Clip-On
1.9
Yes
Free field speakers
0.6+0.9
Yes
Patient response switch
2.0
Yes
USB cable
1.9
Yes
Portable and mobile RF communications equipment can affect the AD528™. Install and operate the
AD528™ according to the EMC information presented in this chapter.
The AD528™ has been tested for EMC emissions and immunity as a standalone instrument. Do not use the
AD528™ adjacent to or stacked with other electronic equipment. If adjacent or stacked use is necessary, the
user should verify normal operation in the configuration.
The use of accessories, transducers and cables other than those specified, with the exception of servicing
parts sold by Interacoustics as replacement parts for internal components, may result in increased
EMISSIONS or decreased IMMUNITY of the device.
Anyone connecting additional equipment is responsible for making sure the system complies with the IEC
60601-1-2 standard.
Guidance and manufacturer’s declaration - electromagnetic emissions
The AD528™ is intended for use in the electromagnetic environment specified below. The
customer or the user of the AD528™ should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The AD528™ uses RF energy only for its internal
function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The AD528™ is suitable for use in all commercial,
industrial, business, and residential environments.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable