EasyManuals Logo

Interface TS14 User Manual

Default Icon
11 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
E-Mail: keiths@interfaceforce.com
Technical changes reserved
TS Series Reaction Torque.doc
Internet: www.interfaceforce.com
Page 11 of 11
9.2 Storage
The storage of the sensors must occur in dry, dust-free rooms, only.
Slightly lubricate shafts and flanges with oil before storing (rust).
10 Disposal
The torque sensors must be disposed according to the valid provisions of law.
11 Calibration
At the time of delivery, torque sensors have been adjusted and tested with traceable calibrated measuring
equipment at factory side. Optionally, a calibration of the sensors can be carried out.
11.1 Proprietary Calibration
Acquisition of measurement points and issuing of a calibration protocol Traceable calibrated measuring
equipment is being used for the calibration. The sensor data are being checked during this calibration.
11.2 DKD-Calibration
The calibration of the sensor is carried out according to the guidelines of the DKD. The surveillance of the
calibrating-laboratory takes place by the DKD. At this calibration, the uncertainty of measurement of the
torque measuring instrument is determined. Further information can be obtained upon request.
11.3 Re-Calibration
The recalibration of the torque sensor should be carried out after 26 months at the latest.
Shorter intervals are appropriate:
Overload of the sensor
After repair
After inappropriate handling
Demand of high-quality standards
Special traceability requirements
12 Data Sheet
See www.interfaceforce.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interface TS14 and is the answer not in the manual?

Interface TS14 Specifications

General IconGeneral
BrandInterface
ModelTS14
CategorySecurity Sensors
LanguageEnglish