4 / 14    P/N 422-6367-ML • REV E • ISS 03DEC18 
Glass type of pane 
inside room 
Double glazed 
with security 
film 
Closing the detector 
To close the detector, follow these steps: 
1.  Insert cover hooks into slots in the base (Figure 7, item 1) 
2.  Close the cover (item 2) 
3.  Tighten the locking screw (item 3) 
4.  Slide down the lock cover (item 4) 
Check the settings with GS960-TR tester 
If the detector is placed too far or too close from the glass, it 
will not respond. At DRC testing the LED of the detector will 
flash 1, 2 or 3 times to show it is in a certain Zone. If it does not 
flash a better place must be selected. 
1.  Put the cover on and make sure it hooks properly into the 
base of the detector.  
2.  Apply power. LED will now indicate your range setting by 
blinking 1 to 3 times. 
3.  Use GS960-TR to test and calibrate the detector for 
optimal position. 
Testing and calibrating 
GS960-TR tester is a specially developed tool for calibrating 
and adjusting the detector GS960 for optimal function in the 
acoustic room – the DRC Digital Room Compensation 
procedure. When testing the detector settings you do not need 
to open it again as the tester will communicate with the 
detector acoustically. Never test the GS960 with a cover 
disassembled from a base. Make sure the cover is fastened 
properly. 
Caution: Do not use the GS960-TR tester in proximity to your 
ears as the tester produces loud noises. 
Digital room compensation procedure 
Prepare the acoustic detector for DRC by setting the D/N line 
(if used) in DAY mode – D/N line should be left open or Low. 
1.  Press the START button of the GS960-TR tester to put the 
power on. Green LED will light. 
2.  Hold the tester 1 to 3 m from the detector and aim the 
speaker at it. 
3.  Press the START button once more to initiate the DRC 
mode. The LED on the detector will start to flicker. See 
Figure 8. 
4.  Go to the furthest distance (max. 9 m) of the glass to be 
protected and aim the speaker at the detector. See 
Figure 9. 
The DRC range calculated by the detector will be 
displayed as a number of pulses from 1 to 3. 
5.  Press the DRC button to send a DRC signal out. Make this 
2 to 10 times from different angles of the protection area 
for optimal capability. The LED will flash confirming it has 
received the signal. The LED will then start to blink and 
flicker. The DRC range calculated by the detector will be 
displayed as a number of pulses from 1 to 3. In case of too 
weak or too strong signals outside the compensation 
range (means that the detector is placed too close or too 
far from the object to be protected), the detector will then 
not show DRC range. 
6.  Press the STOP button from a distance of 1 to 3 m from 
the detector to terminate the DRC procedure. 
If DRC range measured by the detector is different from actual 
DIP switch settings, the LED will continue to blink 1 to 3 times 
showing the correct range number to be set in the detector. 
•  The LED blinks 1 time: set to Zone 1 (4 to 9 m) 
•  The LED blinks 2 times: set to Zone 2 (2 to 4 m) 
•  The LED blinks 3 times: set to Zone 3 (1 to 2 m) 
Protecting several windows by one detector 
Detector can protect several windows in one room if the 
windows are within the coverage area. Make independent DRC 
test for each window to be protected. Follow the procedure 
above for each window. The detector shall be set to the lowest 
Zone number that is the longest range from the detector. 
Timeout 
Both the GS960 detector and GS960-TR tester are equipped 
with a timeout feature. The GS960 will stop the DRC mode and 
the GS960-TR will switch off power if no activity has happened 
within 3 to 4 minutes. 
D/N control (Day/Night) 
The D/N makes it possible to remote control the alarm 
indications of the detectors and remote reset during DAY-
>NIGHT transition. The D/N increases the security of the 
detector, as it enables the alarm indications to be concealed in 
NIGHT mode without any influence on the relay function.  
Self-test 
Detector is continuously checking all vital parts responsible for 
proper detection of glass break signal using internal self-test 
procedure.