Invacare®Perfecto₂™Series
9Troubleshooting
9.1Troubleshooting
SYMPTOMPROBABLECAUSESOLUTION
Alarm:MainPowerLoss:
ShortBeeps,longpause
Concentratornotoperating,
powerswitchOn.
Beep...Beep...
1.Powercordnotpluggedin.
2.Nopoweratoutlet.
3.Trippedcircuitbreaker.
1.Insertplugintooutlet.
2.Inspecthousecircuitbreakersorfuses.Ifproblemrecurs,
useadifferentoutlet.
3.Push/resetcircuitbreaker.Ifproblemrecurs,callservice
provider.
Alarm:
SystemFailure:
Continuous
Concentratornotoperating,
powerswitchOn.
Beep...
1.Unitoverheatingduetoblocked
airintake.
2.Insufficientpoweratoutlet.
3.Internalrepairsrequired.
1.Dooneorbothofthefollowing:
a.Removeandcleancabinetfilters.
b.Moveoxygenconcentratoratleast12in(30,5cm)
awayfromwalls,draperiesorfurniture.
2.DONOTuseextensioncords.Movetoanotherelectrical
outletorcircuit.
3.Callserviceprovider.
281148074-F-06