EasyManua.ls Logo

Invacare IRC5P - Page 32

Invacare IRC5P
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Invacare®Perfecto₂™Series
OperatingAmbientTemperature:Perfecto
2
ModelsandPerfecto
2
VModels–50°F-95°F(10°C-35°C)
at20-60%relativehumidity
Perfecto
2
WModels–50°F-90°F(10°C-32°C)at20-60%relative
humidity
Cabinet:ImpactResistantflame-retardantplasticcabinetthatconformstoUL
94V–0
StandardsandRegulatoryListing:DoubleInsulatedProductETLcertifiedcomplyingwithUL1431and
UL1097ETLcertifiedtoCSAC22.2No.68.
Electrical:Noextensioncords
Placement:
Nocloserthan12in(30,5cm)fromanywall,furniture,draperies,
furnitureorsimilarsurfacestoassuresufficientairflow.
Avoiddeeppilecarpetsandheaters,radiatorsorhotairregisters.
Floorlocationonly.
Noconfinedspaces(Example:Noclosets).
Tubing:
2m(7ft)cannulawithamaximum15m(50ft)ofCrush-ProofTubing
(DONOTpinch)
RelativeHumidity:
20to60%
TimeofOperation:Upto24hoursperday
RecommendedStorageandShippingTemperature:-20°Fto150°F(-29°Cto65°C)at15-95%relativehumidity
321148074-F-06

Table of Contents