SECTION 4—TYPICAL PRODUCT PARAMETERS
Part No 1143482 13 Perfecto
2
Series
NOTE: Stated concentration levels achieved after initial warm-up
period (approximately 30 minutes).
Safety System: Current overload or line surge shutdown. High
temperature compressor shutdown. High
Pressure Alarm w/compressor shutdown. Low
Pressure Alarm w/compressor shutdown.
Battery Free Power Loss Alarm. SensO
2
Oxygen System Possible Obstruction Alert.
Width: 15 in ± 3/8 in (38.1 cm ± 1 cm)
Height: 23 in ± 3/8 in (58.4 cm ± 1 cm)
Depth: 12 in ± 3/8 in (30.5 cm ± 1 cm)
Weight: 43 lbs ± 2 lbs (19.5 kg ± 1 kg) Perfecto
2
Models
45 lbs ± 2 lbs (20.5 kg ± 1 kg) Perfecto
2
W
Models
40 lbs ± 2 lbs (18.1 kg ± 1 kg) Perfecto
2
V Model
Shipping Weight: 48 lbs ± 2 lbs (21.8 kg ± 1 kg) Perfecto
2
Models
50 lbs ± 2 lbs (22.8 kg ± 1 kg) Perfecto
2
W
Models
45 lbs ± 2 lbs (20.4 kg ± 1 kg) Perfecto
2
V Model
Operating Ambient
Temperature:
50°F - 95°F (10°C - 35°C) at 20-60% relative
humidity Perfecto
2
Models and Perfecto
2
V
Model
50°F - 90°F (10°C - 32°C) at 20-60% relative
humidity Perfecto
2
W Models
Exhaust: Less than Ambient + 35°F (+19°C)
Oxygen Output Temperature: Less than Ambient + 6°F (+3°C)
Cabinet: Impact Resistant flame-retardant plastic cabinet
that conforms to UL94V-0.
Regulatory Listing: Double Insulated Product ETL certified
complying with UL1431 and UL1097 ETL
certified to CSA C22.2 No. 68.
Electrical: No extension cords.
Placement: No closer than twelve inches from any wall,
furniture, draperies, or similar surfaces.
Tubing: 7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Recommended Storage and
Shipping Temperature:
-20°F to 150°F (-29°C to 65°C), at 15-95%
relative humidity.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 IRC5PO2 and is the answer not in the manual?

Invacare Perfecto2 IRC5PO2 Specifications

General IconGeneral
Flow Rate0.5 to 5 LPM
Outlet Pressure5.5 PSI
Oxygen Concentration87% to 95.6%
Dimensions23 in (H) x 13 in (W) x 12 in (D)
Electrical Requirements120 VAC, 60 Hz
Operating Temperature55°F to 90°F (12°C to 32°C)
Operating Humidityup to 95% non-condensing
Sound Level≤45 dB(A)
Noise Level≤45 dB(A)