SECTION 6—MAINTENANCE
Perfecto
2
Series 24 Part No 1143482
FIGURE 6.2 Preventive Maintenance Record
MODEL NO. IR C SERIAL NO.
PREVENTIVE MAINTENANCE RECORD
Record Date of Service
Record Elapsed Hours On Hour Meter
Clean Cabinet Filter(s)
Check Prescribed L/min. Flow Rate
- EVERY
4,380 HOURS
Check Oxygen Concentration
UNITS WITHOUT SensO
2
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
Check Power Loss Alarm
Clean/Replace Cabinet Filter(s)
Check Outlet HEPA Filter*
Check Compressor Inlet Filter*
Check Power Loss Alarm
*NOTE: Refer to Preventive Maintenance section of Service Manual.
NOTE:
4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months.
26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years.
Check Oxygen Concentration
UNITS WITH SensO
2
ON EACH INSPECTION
DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS
- EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 IRC5PO2 and is the answer not in the manual?

Invacare Perfecto2 IRC5PO2 Specifications

General IconGeneral
Flow Rate0.5 to 5 LPM
Outlet Pressure5.5 PSI
Oxygen Concentration87% to 95.6%
Dimensions23 in (H) x 13 in (W) x 12 in (D)
Electrical Requirements120 VAC, 60 Hz
Operating Temperature55°F to 90°F (12°C to 32°C)
Operating Humidityup to 95% non-condensing
Sound Level≤45 dB(A)
Noise Level≤45 dB(A)