SECTION 2—INSTALLATION/SEQUENCE OF OPERATION
Perfecto
2
V 12 Part No 1164934
SECTION 2—INSTALLATION/
SEQUENCE OF OPERATION
Verification of Battery Free Power Loss Alarm
CheckthePerfecto
2
Vconcentratorforproperoperatingconditions.
1. Iftheunithasbeeninbelowfreezingtemperatures,allowittowarmuptoroom
temperaturebeforeoperating.
2. Theconcentratormayneedtobeturnedonforfourtofivesecondstochargethe
BatteryFreePowerLossAlarm.Connectpowercordtooutletandturnthe
concentratoron.Turnflowcontrolknobcounterclockwiseandflowwillbegin
IMMEDIATELY.SetflowratetofiveL/min.Lettheunitrunfor30minutes,thenturn
unitoff.
3. UnplugthepowercordandpressOn/Off( )switchtotheOn()position.An
intermittentaudiblealarmwillsound.ThisconfirmsproperoperationoftheBattery
FreePowerLossAlarm.TurnOn/OffswitchOff( ).
4. Connectpowercordtooutletandturnonconcentrator.Unitwillbeeponstartup.
5. Checktheoxygenconcentrationperspecificationsafter3040minutesrunningtime.
Sequence of Operation
TurningOn()thepowerswitchapplies120VACtothecompressormotor,hourmeter,
transformer,coolingfanandtheprintedcircuit(P.C.)board.
Roomairentersthecompressorviathecabinetfilterandthecompressorinletfilter.The
airiscompressedbythewobblepistonsinthecompressortoa
pressureof21p.s.i.(144.79
kPa).
Asincreasedpressurecreatesincreasedtemperature,aheatexchangerisusedtolower
thetemperaturebeforetheairentersthe4wayvalve.Itisthenchanneledtoasievebed
containingtheadsorptionmaterial.Restrictiondownstreamofthesievebedcauses
pressuretobuildupinsidethesievebedwhichisnecessaryfortheadsorptionprocess.A
smallamountofrelativelypureoxygenentersthetopofthesecondbedthrougha
restrictiveleakinthepressureequalization(P.E.)valvewiththebalanceenteringastorage
tank.Thenitrogenremovedisexhaustedbackfromthebedthroughthe4wayvalveinto
roomair.Amufflerislocatedattheexhaustendofthevalvetomufflethesoundofthe
exhaustasitexitstheconcentrator.
Theoxygennotbeingusedtoexhaustischanneledintothestoragetank.Thepressurized
oxygenisregulated
downtofivep.s.i.(34.4kPa),entersanaccurateflowmeasuring
device,flowsthroughOutletHEPAfilterandcheckvalve,thenouttothepatient.
I/O
I/
/O
I/

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 V IRC5PO2V and is the answer not in the manual?

Invacare Perfecto2 V IRC5PO2V Specifications

General IconGeneral
BrandInvacare
ModelPerfecto2 V IRC5PO2V
CategoryMedical Equipment
LanguageEnglish

Related product manuals