TYPICAL PRODUCT PARAMETERS
Part No 1164934 9 Perfecto
2
V
*NOTE:Statedconcentrationlevelsachievedafterinitialwarmupperiod(approximately30
minutes).
IRC5PO2V INDICATORS
Regulatory Listing: Double Insulated Product ETL certified complying with UL 1097
ETL certified to CSA C22.2 No. 68.
Electrical: No extension cords.
Placement: No closer than three inches from any wall, furniture, draperies,
or similar surfaces.
Tubing: 7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO
NOT pinch).
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Recommended Storage and Shipping Temper-
ature:
-20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity.
LABEL
SYMBOL
O
2
PURITY INDICATOR LIGHTS
(LED)
SYSTEM OKAY
O
2
over 85% (± 2%)
GREEN Indicator Light
O
2
Between 73% (± 3%)
to 85% (±2%)
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73% (±3%)
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down.
O
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Perfecto2 V IRC5PO2V and is the answer not in the manual?

Invacare Perfecto2 V IRC5PO2V Specifications

General IconGeneral
BrandInvacare
ModelPerfecto2 V IRC5PO2V
CategoryMedical Equipment
LanguageEnglish