Invacare®SoftformActive®2withRXpump
6Maintenance
6.1Inspection
Itisrecommendedtocheckmattresses(aircellsandcover)
forstrike-through(thismayincludeuidingress,stains,rips
ordamage)afterthereleaseofeachpatientoraftereach
periodofusebyasuitablyqualiedandcompetentperson.
Checkmattresses
1.Unzipthecovercompletely.
2.Checkforanystainingonthewhiteundersideofthe
cover .
3.Checkforanystainingontheinterioraircells.
4.Replaceanystaineditemsanddisposeofasperlocal
authorityprocedure.
6.2CleaningandDisinfection
6.2.1GeneralSafetyInformation
CAUTION!
RiskofContamination
–Takeprecautionsforyourselfanduse
appropriateprotectiveequipment.
CAUTION!
Riskofelectricshockandproductdamage
–Switchoffthedeviceanddisconnectfrom
mains,ifapplicable.
–Whencleaningelectroniccomponentsconsider
theirprotectionclassregardingwateringress.
–Makesurethatnowatersplashestotheplug
orthewalloutlet.
–Donottouchthepowersocketwithwethands.
IMPORTANT!
Wronguidsormethodscanharmordamage
theproduct.
–Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand
mustprotectthematerialstheyareusedto
clean.
–Neverusecorrosiveuids(alkalines,acidetc.),
abrasivecleaningagentsorsolvents(cellulose
thinner,acetoneetc.).Werecommendamild
detergent.
–Alwaysmakesurethattheproductiscompletely
driedbeforetakingitintouseagain.
Forcleaninganddisinfectioninclinicalorlong-term
careenvironments,followyourin-houseprocedures.
6.2.2CleaningIntervals
IMPORTANT!
Regularcleaninganddisinfectionenhances
smoothoperation,increasestheservicelifeand
preventscontamination.
Cleananddisinfecttheproduct
–regularlywhileitisinuse,
–beforeandafteranyserviceprocedure,
–whenithasbeenincontactwithanybody
uids,
–beforeusingitforanewuser.
IMPORTANT!
–Keepacleaningrecordaspartofcleaningthe
system.
6.2.3CleaningInstructions
IMPORTANT!
–Theproductdoesnottoleratecleaningin
automaticwashingplants,withhigh-pressure
cleaningequipmentorsteam.
Acleaningrecordshouldbekeptaspartofcleaning
thesystem.
Cleaningcontrolunit
1.Wipedownthecontrolunitcasingandhosettingswith
adampclothandsuitabledetergent.
2.Usinganylonbrush,gentlycleanallcrevicesasthey
canharbormicroorganisms.
3.Wipeoffthecontrolunitcasingandhosettingswitha
dampclothtoremovealldetergent.
4.Airdryalltreatedsurfaces.
Cleaninghangers
1.Wipedowntheexteriorofthehangersperiodically,
usingadampclothandsuitabledetergent.
Cleaningcovers
(Removalofcontaminantssuchasdustandorganicmatter)
1.Removeallcoversforlaundering.
2.Launderthecoverswithmaximumtemperatureas
statedonproductlabel,usingadiluteddetergent
solution(Instructionsonlabel).
IMPORTANT!
Washingathighertemperatureswillcause
shrinkage.
Dryingcovers
1.Hangcoversfromalineorbaranddripdryinaclean
indoorenvironmentortumbledryonalowheatsetting.
IMPORTANT!
–Tumbledrysettingmustnotexceed40°C.
–Donottumbledryforlongerthan10minutes.
–Drythoroughlybeforeuse.