InvacarSoftformActive®2withRXpump
6Maintenance
6.1Inspection
Itisrecommendedtocheckmattresses(aircellsandcover)
forstrike-through(thismayincludeuidingress,stains,rips
ordamage)afterthereleaseofeachpatientoraftereach
periodofusebyasuitablyqualiedandcompetentperson.
Checkmattresses
1.Unzipthecovercompletely.
2.Checkforanystainingonthewhiteundersideofthe
cover .
3.Checkforanystainingontheinterioraircells.
4.Replaceanystaineditemsanddisposeofasperlocal
authorityprocedure.
6.2CleaningandDisinfection
6.2.1GeneralSafetyInformation
CAUTION!
RiskofContamination
Takeprecautionsforyourselfanduse
appropriateprotectiveequipment.
CAUTION!
Riskofelectricshockandproductdamage
Switchoffthedeviceanddisconnectfrom
mains,ifapplicable.
Whencleaningelectroniccomponentsconsider
theirprotectionclassregardingwateringress.
Makesurethatnowatersplashestotheplug
orthewalloutlet.
Donottouchthepowersocketwithwethands.
IMPORTANT!
Wronguidsormethodscanharmordamage
theproduct.
Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand
mustprotectthematerialstheyareusedto
clean.
Neverusecorrosiveuids(alkalines,acidetc.),
abrasivecleaningagentsorsolvents(cellulose
thinner,acetoneetc.).Werecommendamild
detergent.
Alwaysmakesurethattheproductiscompletely
driedbeforetakingitintouseagain.
Forcleaninganddisinfectioninclinicalorlong-term
careenvironments,followyourin-houseprocedures.
6.2.2CleaningIntervals
IMPORTANT!
Regularcleaninganddisinfectionenhances
smoothoperation,increasestheservicelifeand
preventscontamination.
Cleananddisinfecttheproduct
regularlywhileitisinuse,
beforeandafteranyserviceprocedure,
whenithasbeenincontactwithanybody
uids,
beforeusingitforanewuser.
IMPORTANT!
Keepacleaningrecordaspartofcleaningthe
system.
6.2.3CleaningInstructions
IMPORTANT!
Theproductdoesnottoleratecleaningin
automaticwashingplants,withhigh-pressure
cleaningequipmentorsteam.
Acleaningrecordshouldbekeptaspartofcleaning
thesystem.
Cleaningcontrolunit
1.Wipedownthecontrolunitcasingandhosettingswith
adampclothandsuitabledetergent.
2.Usinganylonbrush,gentlycleanallcrevicesasthey
canharbormicroorganisms.
3.Wipeoffthecontrolunitcasingandhosettingswitha
dampclothtoremovealldetergent.
4.Airdryalltreatedsurfaces.
Cleaninghangers
1.Wipedowntheexteriorofthehangersperiodically,
usingadampclothandsuitabledetergent.
Cleaningcovers
(Removalofcontaminantssuchasdustandorganicmatter)
1.Removeallcoversforlaundering.
2.Launderthecoverswithmaximumtemperatureas
statedonproductlabel,usingadiluteddetergent
solution(Instructionsonlabel).
IMPORTANT!
Washingathighertemperatureswillcause
shrinkage.
Dryingcovers
1.Hangcoversfromalineorbaranddripdryinaclean
indoorenvironmentortumbledryonalowheatsetting.
IMPORTANT!
Tumbledrysettingmustnotexceed40°C.
Donottumbledryforlongerthan10minutes.
Drythoroughlybeforeuse.
12
1627707-D

Other manuals for Invacare Softform Active 2 RX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Invacare Softform Active 2 RX and is the answer not in the manual?

Invacare Softform Active 2 RX Specifications

General IconGeneral
BrandInvacare
ModelSoftform Active 2 RX
CategoryIndoor Furnishing
LanguageEnglish

Related product manuals