6Maintenance
6.1Inspection
Itisrecommendedtocheckmattresses(aircellsandcover)
forstrike-through(thismayincludeuidingress,stains,rips
ordamage)afterthereleaseofeachpatientoraftereach
periodofusebyasuitablyqualiedandcompetentperson.
Checkmattresses
1.Unzipthecovercompletely.
2.Checkforanystainingonthewhiteundersideofthe
cover .
3.Checkforanystainingontheinterioraircells.
4.Replaceanystaineditemsanddisposeofasperlocal
authorityprocedure.
6.2CleaningandDisinfection
6.2.1GeneralSafetyInformation
CAUTION!
RiskofContamination
–Takeprecautionsforyourselfanduse
appropriateprotectiveequipment.
CAUTION!
Riskofelectricshockandproductdamage
–Switchoffthedeviceanddisconnectfrom
mains,ifapplicable.
–Whencleaningelectroniccomponentsconsider
theirprotectionclassregardingwateringress.
–Makesurethatnowatersplashestotheplug
orthewalloutlet.
–Donottouchthepowersocketwithwethands.
IMPORTANT!
Wronguidsormethodscanharmordamage
theproduct.
–Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand
mustprotectthematerialstheyareusedto
clean.
–Neverusecorrosiveuids(alkalines,acidetc.),
abrasivecleaningagentsorsolvents(cellulose
thinner,acetoneetc.).Werecommendamild
detergent.
–Alwaysmakesurethattheproductiscompletely
driedbeforetakingitintouseagain.
Forcleaninganddisinfectioninclinicalorlong-term
careenvironments,followyourin-houseprocedures.
6.2.2CleaningIntervals
IMPORTANT!
Regularcleaninganddisinfectionenhances
smoothoperation,increasestheservicelifeand
preventscontamination.
Cleananddisinfecttheproduct
–regularlywhileitisinuse,
–beforeandafteranyserviceprocedure,
–whenithasbeenincontactwithanybody
uids,
–beforeusingitforanewuser.
IMPORTANT!
–Keepacleaningrecordaspartofcleaningthe
system.
6.2.3CleaningInstructions
IMPORTANT!
–Theproductdoesnottoleratecleaningin
automaticwashingplants,withhigh-pressure
cleaningequipmentorsteam.
Acleaningrecordshouldbekeptaspartofcleaning
thesystem.
Cleaningcontrolunit
1.Wipedownthecontrolunitcasingandhosettingswith
adampclothandsuitabledetergent.
2.Usinganylonbrush,gentlycleanallcrevicesasthey
canharbormicroorganisms.
3.Wipeoffthecontrolunitcasingandhosettingswitha
dampclothtoremovealldetergent.
4.Airdryalltreatedsurfaces.
Cleaninghangers
1.Wipedowntheexteriorofthehangersperiodically,
usingadampclothandsuitabledetergent.
Cleaningcovers
(Removalofcontaminantssuchasdustandorganicmatter)
1.Removeallcoversforlaundering.
2.Launderthecoverswithmaximumtemperatureas
statedonproductlabel,usingadiluteddetergent
solution(Instructionsonlabel).
IMPORTANT!
Washingathighertemperatureswillcause
shrinkage.
Dryingcovers
1.Hangcoversfromalineorbaranddripdryinaclean
indoorenvironmentortumbledryonalowheatsetting.
IMPORTANT!
–Tumbledrysettingmustnotexceed40°C.
–Donottumbledryforlongerthan10minutes.