Why is my Invacare Indoor Furnishing mattress not inflating properly?
- NNancy RamseyAug 6, 2025
If the mattress of your Invacare Indoor Furnishing is not inflating or alternating properly, and you see an alternating system/timing failure indication, begin by checking the CPR hose connection and ensure it is securely locked in place. Inspect the CPR hose and tube connectors for kinks or splits; unkink them or replace if split. If there's a major leak in an air cell or the entire air insert, replace the leaking air cell. If the control unit still doesn't operate after checking the power cable and fuse, or if there's no air due to control unit failure or rotor failure, send the control unit back to Invacare for repair.