EasyManua.ls Logo

Invacare Softform Active 2 RX - 3.5 Symbols on the product

Invacare Softform Active 2 RX
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60129676-A 7
Invacare
®
Softform Active
®
2 Rx
4 Service | Digital Control Unit
4.1 Replacing the Front Housing
R
efers to spare parts kit SP1633237
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
PH1 Phillips key
14 mm Open-ended wrench
1.
Switch o the control unit and disconnect from
mains supply.
2. Turn the unit upside down.
3. Loosen and remove the four screws A.
4. Remove the bottom housing.
5.
Loosen and remove the two screws B fixing the
motor to the front housing.
6. Loosen and remove the three screws C fixing
the compressor to the front housing.
A
A
A
A
Fig. 4-1
B
B
Fig. 4-2
C C
C
Fig. 4-3

Other manuals for Invacare Softform Active 2 RX