EasyManua.ls Logo

Invacare Softform Active 2 - System Installation and CPR; Installing the Invacare Softform Active 2; CPR Procedure Activation

Invacare Softform Active 2
3 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Invacare UKOpera ons Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ, UK
60132743-A 2022-01-10 2 / 3
Installing the system
K J
G
H
I
C D E F
A
B
QuickGuide
Invacare® Softform Active® 2 with RX pump
1.Hangthecontrolunitbymeansofthe2builtinhangers
Aattheendofthebedorplaceitonahorizontal
surface.(Placingthecontrolunitontheoorwillnot
affecttheperformance,butmayexposeittoaccidental
damage.)
2.ConnecttheCPRhoseEtoairoutletsonthemattress
andcontrolunitB,ensuringthatthehoseisnotkinked
ortwisted.Allowspaceforbedrailstodropfreely.
3.Connectthemainspowercabletothepowersupply
socketCofthecontrolunitandasuitableoutlet.
4.Switchonmainspoweroftheoutletifpresent.
5.MovethemechanicalswitchDontheleftsideofthe
ControlunittotheOnposition.Afaintsinglebeep
soundsandthesystemincyclelightasheswhilethe
systempowersup.
CPR procedure
The Invacare® Softform Active® 2 Rx mattress has been fully
tested to comply with the current CPR standard of 5 - 6 cm
compression depth. This was achieved at all stages of
inflation/deflation.
1.HolddownRedCPRbuttonA.
2.Pullhoseconnectorrmlya wayfromthecontrolunitB.
3.Switchoffthecontrolunit.
Mattresswillstarttodeate.Thedeationtimeis
20seconds.
4.WhenCPRiscompletereactivatethesystemfollowing
section Installing the system.
InvacarePortugal,Lda
RuaEstradaVelha949
4465-784LeçadoBalio
Portugal
Invacare Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
UK