M720 pH/ORP Controller Owners Manual
2. Inadequate Flow: Check ow through ow cell and
controller.
3. Blown output fuse in the controller: Replace fuse.
G. Flow LED o
1. Verify that all appropriate valves are open.
2. Verify that there is sucient pressure in the line. Close right
side valve slightly if necessary.
3. Verify that the ow switch is securely connected to the
controller terminals.
4. Both the pH and ORP dose outputs are disabled if the green
Flow LED is not illuminated.