Iradion RS232 Com Test Software
The Iradion Infinity Series Lasers allow user communication to the control card through a utility
software package. This package is documented in 91-3005 LCC com protocol (Customer
Ver). Certain production functions are omitted in the “customer” version because irreversible
changes could be made that would render the laser inoperable.
The software functions on Windows XP, Vista and System 7 operating systems. A properly
configured com port must be available and the com port number must be known before launching
the software.
The available functions are:
Get Status
Get Options
Get Laser Serial Number
Set Auto Fan Enable (if this hardware option is available)
Set Auto Fan Disable (if this hardware option is available)
Set Mod Low Grounded
Set Mod Low Not Grounded
Set Keylock Switch Disable (not available in current software)
Set Keylock Switch Enable (not available in current software)
Set Simmer Frequency (5KHz to 200KHz)
Set Simmer Duty Cycle (0% to 20%) (factory default is 0%)
Set Pulse Stretch (0 µS to 10µS) (factory default is 0 µS)
If the laser control card is reprogramed to a newer version, all of the optional settings must be
manually reset using the Com Test Software, and the correct serial number must also be
installed. A special procedure is required to restore these settings, which is not covered in this
manual.