Functions and Characteristics 
  Copyright ©  Itech Electronic Co., Ltd.  30 
4.  Operate    key and move to OFF. Press [Enter] key (In this case, 
OFF changes to ON, and the  VF  status indicator Trig lamp of the VFD 
display is illuminated). Press [ESC] key to exit setting. 
LIST    OFF        CALL        EDI 
5.  Press [On/Off] key to open input and press [Shift] +    key (Trigger key). 
L1.    0. 0 
Trig 
3.10 Test Function 
The  transient  test  allows  switching  between  two  different  load  values.  A 
common application is to test the dynamic characteristics of DC source. 
There  are  three  different  types  of  transient  operation:  continuous,  pulse, 
toggled. 
Transient test parameters include: 
  Transient test mode 
  Transiently switched A and B values 
  Pulse width time (In PULSE mode) 
  Frequency 
  Duty cycle 
  Current rise and fall slope (CC mode specific parameters) 
3.10.1 Transient Test Function 
Continuous Mode 
In continuous transient operation, the load is continuously switched between 
two load values. An example is shown in the following figure: 
 
 
Take  CC  mode  as  an  example  (other  modes  operate  similarly).  When  the 
measured instrument output voltage is 10V, current is 3A, and load current is 
switched between 1A and 2A, the transient test parameters and steps are set 
as follows: 
1.  Press [Shift]+[2] (Tran), move    key to select ON, press [Enter] to 
confirm. 
TRAN        ON        OFF 
2.  Press    to select transient operation mode as CONTINUOUS (the 
indicator lamp Trig will be lighted). 
MODE    CONTINUOUS   
PULSE    TOGGLE