Automatic Test Function 
  Copyright ©  Itech Electronic Co., Ltd.  47 
The first step: CC mode, current 2A, the upper limit voltage value is 10V, and 
the lower limit voltage value is 2V. 
1.  Press [CC], set the current to 2A, and then press [Shift]+[CV] (Setup) to 
enter the setting interface. 
RANGE=30.000A 
CC 
2.  Set the maximum working current, and press [Enter] to confirm. 
RANGE =2.000A     
CC 
3.  Set the upper limit of voltage, press [Enter] to confirm. 
HIGH=10.00V 
CC 
4.  Set the lower limit of voltage, press [Enter] to confirm. 
LOW=2.000V 
CC 
5.  Set the rising slope of current, press [Enter] to confirm. 
UP=1A/uS 
CC 
6.  Set the descending slope of current, press [Enter] to confirm. 
DOWN=1A/uS 
CC 
The parameter settings are complete. 
10.0000V      0.000A 
0.00W          I=2.000A 
7.  Press  [Shift]  +  [4]  (Save)  to  enter  the  save  settings  interface.  Press 
numeric key [1] twice to save the current step as the first step of Program 
2. 
SAVE    11 
The second step: CV mode, voltage 3V, the upper limit current value is 5A, and 
the lower limit current value is 0A. 
1.  Press [CV], set the voltage to 3V, and then press [Shift]+[CV] (Setup) to 
enter the setting interface. 
RANGE=120.00V 
2.  Set the maximum working voltage, take 3V as example, and press [Enter] 
to confirm. 
RANGE=3.00V 
3.  Set the upper limit of current, press [Enter] to confirm. 
HIGH=5.000A 
4.  Set the lower limit of current, press [Enter] to confirm. 
LOW=0.0000A 
The parameter settings are complete. 
10.0000V      0.000A 
0.00W            V=3V 
5.  Press  [Shift]  +  [4]  (Save)  to  enter  the  save  settings  interface.  Press 
numeric keys [1] and [2] to save the current step as the second step of 
Program 2.