Communication Interfaces 
  Copyright ©  Itech Electronic Co., Ltd.  69 
  RS-232 Interface 
In  RS-232  interface,  all  SCPI  commands  can  be  used  for  programming.  If 
RS-232 interface is selected, in accordance with internal connection of data 
terminal  equipment  (DTE)  and  data  communication  equipment  (DCE)  as 
defined in EIA RS-232, the load is connected to another DTE (e.g., PC COM 
interface) with direct-connected Modem cable. 
 
  Communication Setup 
Please ensure the PC and the load have the same configuration in the following 
items. 
Baudrate: 9600(4800, 9600, 19200, 38400). You could enter the system menu 
to set the baudrate. 
Data bit: 8 
Stop bit: 1 
Parity bit: (none, even, odd) 
EVEN    8 data bits have even parity
 
ODD  8 data bits have odd parity 
NONE  8 data bits have no parity
 
Native machine address: (0 ~31, factory default is 0)   
 
USB Interface 
Connect  the  load  and  the  computer  using  a  cable  with  two  USB  interfaces 
(each end). All functions of the load can be programmed via USB. 
After connecting the load and computer by USB, you need to install IT-E122 
driver (download from 
http://www.itech.sh/support/Down-3.htm?model=IT-E122%20%E5%85%89%E
9%9A%94%E7%A6%BB%E9%80%9A%E8%AE%AF%E6%8E%A5%E5%8F
%A3%E9%A9%B1%E5%8A%A8 or contact ITECH agent). The device 
manager of PC will display ‘Prolific USB-to-Serial COM Port’ after installing. 
Notes: Only  IT8513B+/IT8514B+/IT8514C+/IT8516C+ models  have  the USB 
communication interface. After the driver is installed successfully, use USB line 
to connect in communication. (Please don’t connect DB9 interface at the same 
time). Don’t need to set the menu.