EasyManua.ls Logo

JBL PROFLORA CO2 CONTROL - Initial Start-up; 3.1 Preparation; 3.2 Language selection; 3.3 Calibration

JBL PROFLORA CO2 CONTROL
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3. Initial start-up
When the device is switched on for the first time, the display language is selected and then the unit needs to be calibrated� It is not possible to call up
other menu items beforehand� This is for your own safety as the device would measure incorrect values without successful calibration� Afterwards,
any settings can be made�
If the value overview (chap� 4) appears instead of „Language selection“, the unit has already been used (demonstration etc�)� In this case calibration
must be carried out� To do this, tap the „Settings“ symbol and select the menu item „Calibration“� Confirm with OK� Continue as described below
3.1 Preparation
Please use the calibration liquids and cuvettes supplied with the JBL PROFLORA CO₂ pH SENSOR SET Fill each calibration cuvette up to the 10 ml
mark, filling one cuvette with pH 7�00 buer solution, one with pH 4�00 buer solution and one with deionised water (JBL PROFLORA CO₂ DEST)
The buer solutions are coloured with an indicator to avoid confusion� For more stability, place the calibration cuvettes in the 3 large holes on the
calibration holder� Unscrew the screw fitting on the storage tube of the sensor by about one turn and pull out the sensor Put the sensor and the
temperature sensor in the calibration cuvette filled with deionised water, sway both sensors a little and leave them there� Since the pH measurement
and the calibration are temperature-dependent, both sensors must always be immersed in the calibration liquid�
3.2 Language selection
Select your desired language by tapping the arrow symbols and confirm with OK� The following choices are available: DE/EN-UK/FR/NL/IT/DK/ES/PT/
RO�
3.3 Calibration
• Tap OK to start the calibration�
• Follow the instructions in the display panel and immerse both sensors into the appropriate buer solution, first in one solution and then the other
The device guides you through a so-called two-point calibration� In this process, the device is first calibrated with buer solution pH 7�00 and
then with buer solution pH 4�00, never the other way round! The temperature is irrelevant because it is measured by the temperature sensor and
temperature influences are automatically compensated�
• After successful calibration, the display shows in this order: „Calibration, pH 7�00 OK“, „Calibration, pH 4�00 OK“�
• Confirm both with OK�
• Tap the settings symbol� The value overview appears, informing you about the most important values (chap� 4)�
• Dispose of the used buer solutions, briefly rinse the cuvettes in tap water and dry them
Install the pH sensor in a very dark place with good water flow in the aquarium� The sensor can be immersed to a maximum of 2/3 of its length� The
sensor cap with cable must not be immersed in the water The suction holders provided allow the temperature sensor to be attached anywhere in the
aquarium� Now you can select your required settings on the menu (chap� 7�1 – 7�13)�
4. Sleep mode and value overview
If no input is made, the device will revert to the sleep mode after about 30 seconds� The image of a water drop and the permanent information of
the current pH and temperature values will appear After tapping the screen the value overview will appear� It gives information, as illustrated below,
about the most important processes relating to the pH value and the supply of CO₂ into the aquarium:
1 Switch field “Settings” (the way to all possible settings)
2 Status line Alarm with adjusted limit values for pH and temperature
3 Status line Hysteresis with adjusted value
4 Actual pH value (current pH value in buer or in your aquarium)
5 Status line Auto pH with adjusted KH
6 Current temperature (illustrated flashing read in the alarm status)
7 Valve status (illustrated in automatic mode with valve on continuity)
8 Time to calibration (days/hours)
9 Sensor voltage in mV
10 Current precision of measurement (re-calculated for each calibration)
11 pH set point (required pH value, adjusted by device through CO₂ supply)
12 Warning triangle (flashes in case of alarm, while the aected value flashing in red)
13 Key lock (unlocked in illustration)
5. pH set point adjustment
If you are using a pH control device for the first time, we strongly recommend that you initially use the auto pH menu point to adjust the pH set point:
Touch the switch field “Settings” and select the menu point “Auto pH” by tapping on the arrow symbols� Confirm with “OK” Enter the current KH
value of your aquarium water by pressing the fields “plus” or “minus” Confirm with “OK” It is important you measure the KH level in your aquarium
by means of a KH test, such as the JBL PROAQUATEST KH
The device will now calculate the optimal pH value for this KH and save it as set point� The calculation is based on a CO₂ content of 22�5 mg/l, which
is considered as optimal�
5.1 Control
When the valve is switched to automatic (factory setting, see chap� 7�7), the device adjusts the following:
Current actual value in the aquarium: pH 7�80� Auto pH has set pH 7�10 as set point� Symbol for valve status shows “open” and “A for automatic�
CO₂ is now added until a pH value of 7�00* is attained� The device then closes the solenoid valve and the symbol for valve status shows “shut” and
“A for automatic�
Consumption by the plants and the compensation from the atmosphere leads to the CO₂ level decreasing and the pH level slowly increasing� After
reaching pH 7�20* the solenoid valve opens again and CO₂ is added again, until the pH value has dropped again to 7�00*, and so on�
*For pre-selected hysteresis (switching point) of 0�1 (factory setting, see chap� 7�6)
UKUK
18 19
1
12
13
2
11
3
10
4
9
5
8
6
7
1
12
13
2
11
3
10
4
9
5
8
6
7

Table of Contents

Related product manuals