EasyManua.ls Logo

JBL PROFLORA CO2 CONTROL - Operational Notes Part 2; 8.2 Return function; 8.3 Sleep mode; 8.4 Cleaning

JBL PROFLORA CO2 CONTROL
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.2 Return function
If no key is pressed for 30 seconds the display automatically returns to the value overview Any values not confirmed by OK will not have been stored�
8.3 Sleep mode
If no key is pressed for 10 minutes, the display backlighting fades to the lowest setting� When you tap the display, it switches back to the previously
set brightness level�
8.4 Cleaning
The surface of the device may be cleaned with a soft, slightly damp cloth� Do not use cleaning agents� Do not immerse the device in water!
8.5 Power failure
In the event of a power failure, all the values previously set will remain stored� Only the time calculation for the calibration reminder is interrupted
during the power failure�
8.6 Use of JBL PROFLORA CO₂ CONTROL in combination with calcium reactors in marine water
Calcium reactors are tasked with dissolving the lime out of the substrate within the reactor (e�g� coral gravel), and returning it as carbonate hardness
into the marine water aquarium� Thus they counteract the consumption of KH in the aquarium and no KH supplements need to be added� The KH of
natural marine water in oceans is between 6 and 9 °dKH, depending on the ocean� CO₂ is the best way to eectively release KH from the substrate in
the calcium reactor This process requires a pH level of approx� 6�5 inside the calcium reactor, which needs continuous monitoring�
For calcium reactors which have an opening for pH sensors, this is very easy:
All you have to do is install the pH sensor there and set the pH set point on the JBL PROFLORA CO₂ CONTROL to 6�5 (see chap� 7�4)� The connecting
cable for the solenoid valve needs to be connected to the solenoid valve of the CO₂ system, which introduces the CO₂ into the calcium reactor
For calcium reactors without an opening for a pH sensor, the sensor has to be installed inside the water stream leaving the calcium reactor, by means
of a suitable device� Settings as described above�
9. Note on handling pH sensors
The pH sensor is the most delicate part of the whole measuring equipment and therefore needs special care when handling� If the following few
handling tips are followed, the sensor will provide accurate measurements for a long time�
9.1 Handling and care
• Avoid hard impacts and take particular care when handling the delicate sensor tip made of special glass�
• Avoid kinks in the sensor cable�
• Never allow the sensor tip to dry out!
• Sensors which are not in use for a longer period of time should always be placed in JBL storage solution�
• Never fully immerse the sensor in the water The sensor cap and cable connection should always be out of the water� Ideally the sensor should be
immersed to the top of the writing on the sensor shaft (JBL pH Sensor)�
• Place the sensor in as dark a place as possible in the aquarium to prevent algae growing on the tip� Algae growth could lead to incorrect measure
-
ments
• Do not lay the sensor cable next to mains current-bearing cables for long distances�
• If the sensor is accidentally pulled out of the water or the water level drops significantly, the device will measure incorrectly and adjust accordingly�
This is dangerous for the fish� Regular checking is therefore recommended�
• Any dirt that has accumulated on the tip of the sensor can be carefully wiped o with a soft cloth� Never rub vigorously, just dab�
9.2 Service life of the sensor
pH sensors are subject to natural wear and tear, also known as ageing� Ageing begins from the day of manufacture� The voltage delivered by the
sensor tip and recorded by the measuring device and transformed into pH units gives an indication of the condition of the sensorThe voltage in mV
(millivolt) can be read directly from the JBL PROFLORA CO₂ CONTROL
When dipped into pH 7�00 buer solution, a new sensor shows a voltage of 0 +/- a few mV This voltage changes per full pH unit by about 59 mV in
a positive or negative direction, depending on whether measurements are above or below 7�00� When a new sensor is dipped into pH 4�00 buer
solution, the voltage is approx� 177 mV If the sensor has aged, voltage measured at pH 7�00 usually shifts into the negative range�
For example, in 7�00 buer solution, -28 mV is measured� In addition, the voltage dierence per full pH unit decreases� For example, for pH 4�00
buer solution, 110 mV can still be indicated, which would correspond to a dierence of 46 mV per pH unit� For pH 7�00 the JBL PROFLORA CO₂
CONTROL accepts a shift of up to 115 mV and a reduction in voltage dierence per full pH unit of up to 35 mV For measurements outside these
limits the sensor is rejected as faulty at the end of the calibration�
The average service life of a pH sensor is 24 months� Depending on handling and care, ageing can be faster or slowerThe ageing process is accele
-
rated by the continual measurement of extreme pH values, neglected calibration, dirt etc
9.3 Handling buer solutions
Each pH measurement is only as good as the calibration carried out� Therefore, in your own interest, please follow the advice below:
• Store the JBL buer solutions in a cool place and out of the reach of children�
• Always take the amount of buer solution required for a calibration procedure fresh from the storage bottle�
• Never re-use old buer solution, always dispose of it at the end of each calibration�
• Never pour used buer solution back into the storage bottle�
10. Technical data
Display Touch 2,8“
pH measuring range pH 3�00 – 9�00� All values outside this range appear as 2�99 or 9�01�
Setting range for pH set point 5,00 – 9,00
Display / pH measurement accuracy 0,01 pH / 0,02 pH*
Temperature measurement range 0,1 – 84 °C
Temperature measurement accuracy 0,06 °C
Temperature compensation automatic
Calibration time max� 1�5 minutes per buer solution
CO₂ content as basis of calculations for auto pH curve 22,5 mg/l
Languages DE/EN-UK/FR/NL/IT/DK/ES/PT/RO
Voltage 12 V DC
Control voltage for solenoid valve 12 V DC
Power input max� 0,5 W
Power supply unit Primary: 100 – 240 V AC, 47 – 60 Hz, 0�25 A
Secondary: 12 V DC, 0�3 A, 3�6 W
* depending on age and condition of sensor
UKUK
24 25

Table of Contents

Related product manuals