EasyManua.ls Logo

JBL PROFLORA CO2 CONTROL - Menu System Configuration Part 2; 7.7 Valve; 7.8 Alarm; 7.9 Brightness

JBL PROFLORA CO2 CONTROL
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.7 Valve
Open with the following steps:
Settings key > select Valve > press OK > select “auto”, “man open” or “man shut” > press OK.
In this menu item you can select the operation mode of the externally controlled solenoid valve� The options are automatic operation (auto), manual
open (man open) and manual shut (man closed)� For the regulation feature to work, the “auto“ setting (factory setting) needs to be selected� For the
settings “man open “ and “man closed“, the valve stays open or shut until the setting is changed again�
A green symbol for “open“ or a red symbol for “shut“ and “A“ for automatic or “m“ for manual appear on the value overview (chap� 4)�
7.8 Alarm
The device has a versatile alarm function for the pH value and temperature�
Open with the following steps:
Settings key > select Alarm > press OK > use selection key to select alarm type (“alarm o ”, “flashing”, “flashing and tone”) > press OK
> select your chosen limit values for pH and temperature with „plus“ and „minus“ > press OK.
pH alarm:
If the current pH value in the aquarium exceeds the set target value upwards or downwards by more than the value entered, a warning triangle will
flash on the display and the pH value indication will change to red and flash (if “flashing” was selected)� Where “flashing and tone” was selected, an
acoustic signal will sound as well�
The alarm limit can be set as follows: +/- 0�10 to +/- 2�00 pH in steps of 0�05 pH� Factory setting: +/- 1�00 pH
Tip: We recommend setting the alarm value for the pH to +/- 0,50 to check the function of the CO₂ supply� If the device then triggers an alarm, check
your CO₂ system to see if the cylinder is empty or the supply amount has been adjusted etc�
Temperature alarm:
The temperature alarm function enables you to set the upper and lower alarm limits separately� Thus you can freely select the alarm limits relevant
to your aquarium� If one of the limit values is exceeded or not reached, the same occurs as for the pH alarm�
Factory setting: 20 °C < t < 30 °C
The limit values set appear in the status line on the value overview (chap� 4)�
7.9 Brightness
Open with the following steps:
Settings key > select Brightness > press OK > select the desired value using “plus” and “minus” > press OK.
Factory setting: middle level�
7.10 Contrast
Open with the following steps:
Settings key > select Contrast > press OK > select the desired value using “plus” and “minus” > press OK.
Factory setting: middle level�
7.11 Key lock
Open with the following steps:
Settings key > select Key Lock > press OK > confirm with “locked“/”unlocked“ > press OK.
For selection “locked“ > enter the password (4-digit number) > press OK.
After locking the 4-space number field appears when any key is pressed� Enter the 4-digit password and confirm with OK� The keys are now unlo
-
cked again
In the event that you have forgotten the password, you can unlock the device with the following master password: 7422�
In the value overview (chap� 4) a symbol for keys locked (red padlock locked) and for keys unlocked (green padlock unlocked) appears�
7.12 Software version
Open with the following steps:
Settings key > select software version > press OK
The version of the software currently installed and the serial number of the device are displayed� This information is only required for service purpo
-
ses
7.13 Reset function
Open with the following steps:
Settings key > select “Restore all factory” > press O.K.
If you press longer than 1 sec on OK, all individual settings will be deleted and reset to the factory setting. The master password
7442 is retained.
Factory setting after reset:
Language: German
Calibration reminder: 30 days
Auto pH: ON, KH 8 / pH 7�03
Hysteresis: 0�10
Valve: auto
Alarm: pH +/- 1,00; 20 °C < t < 30 °C
Brightness: medium value
Contrast: medium value
Key lock: unlocked
Master password: 7422
8. Miscellaneous
8.1 Calibration
Display during waiting time: Whilst the sensor is in the buer solutions the device measures the sensor voltage delivered and waits until the signal
has stabilised� This takes a maximum of 1�5 minutes per buer solution�
The following appears on the display during the waiting time:
First line: pH value of the required buer solution and voltage in millivolt
Second line: temperature in °C
Third line: progress bar
The far right of the display also shows the current accuracy of the measurement, which is redefined for the sensor used during each calibration�
Example: +/- 0�02 pH� The higher this value is, the worse the condition of the sensor is�
Calibration reminder: the device has an automatic calibration reminder function which gives you a reminder for calibration every 30, 45 or 60
days� The time to the next calibration is displayed in days and hours in the value overview (chap� 4)� The remaining time display also appears on the
top right of the idle screen 5 days before the calibration period expires� Its colour changes to red two days before calibration is due� After the expiry
of the deadline the display flashes and counts the days with a minus sign, to remind you how many days the calibration is overdue for
If calibration cannot be carried out at that moment, the device continues to measure and control as before� However, the warning signal (flashing)
can only be cancelled by calibration�
The time to the next calibration can only be recorded as long as the device is connected to the power supply� After a longer pause, it is vital that
calibration is carried out before the device is again used to measure and regulate�
Error messages during or after the calibration:
The measuring behaviour of pH sensors changes with age i�e� it begins to deviate from the original setting� Regular re-calibration of the device (com
-
parison with the changed measuring behaviour of the sensors) is vital to obtain reliable results
As a rule, the more often, the better
If, at the end of the calibration process, the symbol of the operating instructions and an error message such as DELTAV, OFFSET or similar appear, the
sensor may be defective� Then repeat the calibration�
In the event that calibration is unsuccessful a second time, consult the following table:
Error message Cause Solution/Action
DELTAV Voltage dierence per full pH unit is less than
35 mV: sensor too old or accidentally calibrated
twice with same buer solution�
Repeat calibration, check correct buer
solution is used� If unsuccessful, connect new
sensor and calibrate�
OFFSET Sensor voltage at pH 7�00 is outside the range:
sensor too old, cable faulty, no sensor connec-
ted or initially calibrated with buer solution
4
�00 by accident�
Check sensor cable for damage, connect
sensor correctly and repeat calibration (starting
with buer solution 7�00!)� If unsuccessful,
connect new sensor and calibrate�
UNSTAB Unstable sensor voltage: sensor tip soiled,
electrolyte leaked out of sensor or sensor tip
broken o�
Clean sensor tip� If unsuccessful, connect new
sensor and calibrate�
UKUK
22 23

Table of Contents

Related product manuals