62 ORL-55 JDSU
8 R
EMOTE CONTROL
Return Loss commands
ORL-55 and cORL-A1 additional
remote commands
Command
string
Parameter type / Response type /
Unit / Info
:SOUR:ID See „:SOUR:ID” on page 51.
:SOUR:ID? See „:SOUR:ID?” on page 51.
:SOUR
:STAT
See „:SOUR:STAT” on page 52.
:SOUR
:STAT?
See „:SOUR:STAT?” on page 52.
:RET
:LOSS
:NORM
Triggers the “Normalizing Open Port”
function in Return Loss mode.
Note: This operation takes a few
seconds to complete.
:FETC
:MEAS
:VAL?
See „:FETC :MEAS :VAL?” on
page 61.
Command
string
Parameter type / Response type /
Unit / Info
:EXEC:REF <reflections factor>
Paramter type: integer
Reflections factor in thousandth: for
example 950 for 95%.
For example for 95% type in
:exec:ref 950