18
Joerns Bariatric Beds
Model Bari10A
© 2015 Joerns Healthcare • 6110089 RevF • 15-2235
Bari10A Bed Troubleshooting
Pendant and Sta Control Troubleshooting
Problem Solution
If movement does not occur when
Pendant or Staff button is pushed
1. Function may be locked. Check Staff Control function lockout.
2. Adjustment may already be at maximum or minimum level.
3. Bed is not plugged into wall power outlet and battery has
discharged. Plug bed into wall outlet.
4. Pendant cord may be damaged. Check cord for damage. If cord is
damaged, discontinue bed usage until cord is replaced.
If adjustment only partially occurs,
stops and beeps
1. Battery may be low. Plug bed into power source.
2. Bed may be overloaded or constrained from moving. Determine load
and check for obstructions.
3. If beeping occurs while raising or lowering bed, Hi-Low actuators
may have lost synchronization. Lower bed to lowest position.
Release Bed Down button. Press and hold Bed Down button for ve
seconds to recalibrate synchronization. If problems persist, contact
Joerns Healthcare Technical Support at 800-966-6662.
Bed not responding to staff control
1. Check lockout status.
2. Check connection is fully inserted with cable secured.
3. Check the bed is plugged into power source and control box green
light is illuminated.
4. Unplug bed for 60 seconds and plug back in.
5. If these steps fail to resolve the problem, call Technical Support
at 800-966-6662.
All lockouts indicators ashing
1. Push and hold Hi-Lo up and down on staff control or pendant
simultaneously until ashing stops (No more then 15 seconds).
2. Unplug bed for 60 seconds and plug back in.
3. If these steps fail to resolve the problem, call Technical Support
at 800-966-6662.
Knee Lockout indicator ashes
when attempting articulation
1. Bed is in a reverse Trendelenburg or Comfort Chair position, level
bed (page 8, Figure 6, button 9).
2. Lower or raise Hi-Lo to end of travel.
3. If these steps fail to resolve the problem, call Technical Support
at 800-966-6662.