3
Joerns Bariatric Beds
Model Bari10A
© 2015 Joerns Healthcare • 6110089 RevF • 15-2235
Warning: Possible Injury Or Death. If using
accessories not manufactured, marketed, or provided
by Joerns Healthcare for Joerns
®
beds, consult with
the manufacturer for compatibility and limitations prior
to use. Failure to do so may result in injury or death.
Warning: Possible Shock Hazard. Bed power
cord has hospital-grade, 3-pronged grounded plug.
Grounding reliability can only be achieved when
equipment is connected to an equivalent receptacle
marked “hospital-grade”. If external grounding
conductor is in doubt, operate bed from backup
battery. Failure to do so could result in injury.
Warning: Possible Equipment Damage Or Injury
To Staff Members. Service and repair must only be
performed by qualied, authorized Joerns Healthcare
Technical Support personnel (call 800-966-6662).
Warning: Explosion Hazard. Do not use in the
presence of ammable anesthetics, gas or vapors.
Failure to heed this warning could result in equipment
damage, injury or death.
Caution: Stay clear of underneath the bed when
adjusting or moving the bed.
Caution: Mattress support platform is not grounded,
nor does it have a potential equalization connection
point. To prevent injury to resident/patient connected
to intravascular or intercardial medical equipment,
alternate measure must be implemented to equalize
the potential between mattress support platform and
medical equipment.
Warning: To reduce the risk of injury, ensure the
mattress support platform is horizontal and in the
lowest position with the siderails fully raised and
locked when moving the bed with a resident/patient
in it. For best maneuverability, push the bed from the
head end.
To avoid injury to the resident/patient and/or user, do
not attempt to move the bed laterally with the steer
engaged. The wheel cannot swivel.
Warning: The Bari10A should not be used adjacent
to, or stacked with, other equipment due to the risk
of electronic interference. If adjacent or stacked use
is necessary, the bed should be observed to verify
normal operation in the conguration in which it will be
used.
Warning: Using non-approved cables and accessories
may negatively affect the EMC performance.
Warning: Mobile RF communication equipment can
affect Medical Electrical Equipment.
Warning: Medical electrical equipment needs special
precautions regarding EMC and needs to be installed
and put into service according to the EMC information
provided in this manual.
Caution: Particular precaution must be considered
during exposure to strong emission source such as
High Frequency surgical equipment and
similar devices. Do not route High Frequency cables
on or near device. If in doubt, contact a qualied
technician.
Warning: Possible Injury or Death. Make sure
mattress support platform is in non-extended position
(narrow) and all casters are locked during resident/
patient transfers.
Caution: To avoid damage to the siderail mechanisms,
do not move the bed using the siderails. Use the push/
pull handles integrated in the end panels to move the
bed.
Caution: Before transporting bed, unplug bed from
power source, wrap Power Cord around cord storage
hooks on bed frame below Head Panel and hang
Pendant from Head Panel. Failure to do so could
result in Power Cord, pendant cord, or bed damage,
thereby creating a potential hazard.
Caution: Drainage bag loops are not intended for use
as tie downs.
Do not position the device where access to mains
connection cannot be disconnected.
There are no known contraindications for this device.
This bed conforms to Application Environment 3 of IEC
60601-2-52.
When assessing the risk for entrapment, you need
to consider your bed, mattress, Head/Foot Panels,
Side Rails/Assist Device and other accessories, as
a system. It is also extremely important to review
the resident/patient’s physical and mental condition
and initiate an appropriate individual care plan to
address entrapment risk. All Joerns bed systems are
evaluated for compliance to the FDA/CDRH “Hospital
Bed System Dimensional and Assessment Guidance
to Reduce Entrapment.” For more information, call
Customer Care, 800-826-0270.