2. MAINTENANCE
JCY-1900
1950
DR Service Manual
47
DOC No. 7ZPNA4409
2.9. Software Update
2.9.1. Software Update
The MCB, ARB and OPU software can be updated. Refer to Installation Manual (Software Part) section 9 for more in-
formation.
1) Log in to the maintenance software, and select [Software Update] ⇒ Tab[SOFTWARE UPDATE].
2) Select Tab[RCU] or [OPU].
3) Click [Browse], and select program file for update.
4) Click [Update], and then update is started.
2.9.2. Setting of expiration date in Replacement Parts
Set the expiration date (month / year) of replacement parts. Refer to Installation Manual (Software Part) 4.3.2 for more
information.
1) Log in to the maintenance software, and select [Home]⇒Tab[Installation]⇒[SYSTEM].
2) Select Tab[Replacement parts].
3) Input the expiration date, and click [Set].
4) Operation time (unit: time) of some parts is calculated and displayed by the VDR. After Replacement, click [Clear]
to clear the value.
(When Expiration passed 1 year, ● (red) is displayed.)
* Check the setup from the OPU after setup. (Refer to 3.4.3.)