ME
FP89-ME
Working temperature range °C -90 ... 100
Temperature stability °C ±0.02
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20 -40 -60 -80
1.0 0.92 0.88 0.75 0.58 0.2
Cooling compressor 2-stage
Refrigerant R-404A + R-508b
Overall dimensions (WxDxH) cm 55x61x90
Bath opening (WxL) cm 13x15
Bath depth cm 16
Filling volume liters 5.5 ... 8
Weight kg 133
Mains power connection V/Hz 230 / 50
Current draw (at 230 V) A 16
Mains power connection V/Hz 230 / 60
Current draw (at 230 V) A 16
All measurements have been carried out at:
rated voltage and frequency, ambient temperature: 20 °C
Technical changes without prior notification reserved.
ME
Temperature selection
digital
via keypad
indication on VFD COMFORT-Display
remote control via personal computer
indication on monitor
Temperature indication
VFD COMFORT-DISPLAY
Resolution
°C
0.01
Absolute Temperature Calibration INT/EXT
°C
±3 / ±9
Temperature control
PID3 cascade temperatur control
Heater wattage (at 230 V)
kW
2,0
Heater wattage (at 115 V)
kW
1,0
Electronically adj. pump capacity
stages
1 ... 4
Flow rate at 0 bar
l/min
11 ... 16
Pressure max. at 0 liter
bar
0.22 ... 0.45
Electrical connections:
External alarm device
Vdc/mA
24-0 / max. 25
Computer interface
RS232
External temperature sensor
Pt100
Ambient temperature
°C
5 ... 40