39
Important times for a command transmission:
To ensure a safe data transfer, the time gap between two commands should
be at least 250 ms.
The recirculating cooler automatically responds to an in command with a data
string followed by a LF (Line Feed). The next command should only be sent
after 10 ms.
The commands are divided into in or out commands.
in commands: asking for parameters to be displayed
out commands: setting parameters
The out commands are valid only in remote control mode.
Examples:
Command to set the working temperature to15,5 °C:
out_sp_00
15.5
Command to ask for the working temperature
in_sp_00
Response from the recirculating cooler:
15.5 LF