ROTARY DIAL POSITIONS
STATUS
Displays the battery level and the number of days until
calibration is due
AUX
Displays up to two pages of two lines that toggle for custom
setting
O
2
/ Eff
Displays calculated Oxygen values and the calculated
efficiency when O
2
values are less than 18%
EFFn, EFFg or EFFc as selected by the user.
CO
2
&CO
Displays Carbon Dioxide in % and Carbon Monoxide in ppm
RATIO
Displays the CO/CO2 ratio and the selected fuel type
TEMP
Displays the flue (Tf) and inlet (Ti) temperatures in
o
F
or
o
C .
If the flue probe’s temperature sensor is broken or open circuit
Tf displays -OC-
o
C
FUEL
SET FUEL, select the boiler/furnace fuel using UP/DOWN and
ENTER keys. Preprogrammed fuels: NAT GAS, Light Oil,
PELLETS(wood), BUTANE, L.P.G., PROPANE
DATE /
TIME
Displays the date and time