4.2 PRESSURE TEST
Select “Prs”. The pump stops automatically. Press
to auto-zero the
pressure sensor. Using the black connectors and manometer hose, connect to P1
for single pressure or P1 and P2 for differential pressure.
PRS display
PRESSURE
Defaults to smoothing off on start-up. Can
be changed via “Menu”.
P 0.01
Defaults to low resolution on start-up. Can
be changed via “Menu”.
mBar
Pressure units can be changed via “Menu”.
14:56:29
Displays time to enable manually timed
tests
Press
to print a pressure test (also sends to PC if Bluetooth fitted).
Hold
for 2+ seconds to log a pressure report.
Viewing / printing a logged pressure test
Select MENU / REPORT / PRESSURE / VIEW
Use
or to select the log number to be viewed / printed.
Press to print the test (also sends to PC if Bluetooth fitted).
SAFETY WARNING
Before using the KANE425 to measure the pressure of a gas/air ratio valve, read
the boiler manufacturer’s instructions thoroughly. If in doubt contact the boiler
manufacturer.
After adjusting a gas/air ratio valve it is essential to confirm that the CO,
CO2 and CO/CO2 ratio readings are within the boiler manufacturer’s
specified limits.