4.5 ROOM CO TESTING
Select “Room CO” for CO investigations.
Press
to start the 15 minute duration room CO test.
ROOM CO display
ROOM
CO
When complete the room CO test is
automatically stored in the memory
CO 00
Real time CO reading (ppm)
TEST 00
Test 00 = initial reading
Test 15 = final reading
LOG 01
The log number for the test is displayed.
Press
to print the room CO test (also sends to PC if Bluetooth fitted)
Viewing / printing a logged Room CO test
Select MENU / REPORT / ROOM CO / VIEW
Use
or to select the log number to be viewed / printed.
Use to view individual readings.
Hold
for 2+ seconds to print the test (also sends to PC if Bluetooth
fitted).
NOTE: To abort a room CO test move the rotary switch to any position
other than “Room CO”.
NOTE: The “AUX” display can be used to time and manually record ambient
CO levels to comply with BS7967 – see Page 14.