8. ANALYSER PROBLEM SOLVING
If any problems are not solved with these solutions, contact us or an authorized
repair center.
Fault symptom
Causes / Solutions
Oxygen too high
CO
2
too low
Air leaking into probe, tubing, water trap,
connectors or internal to analyser.
Oxygen cell needs replacing.
Oxygen reading (- - - -)
CO reading (- - - -)
Analyser was stored in a cold environment
and is not at normal working temperature.
Oxygen cell or CO sensor needs replacing.
Batteries not holding
charge
Analyser not running on
mains adapter.
Batteries exhausted.
AC charger not giving correct output.
Fuse blown in charger plug.
Analyser does not respond
to flue gas
Particle filter blocked.
Probe or tubing blocked.
Pump not working or damaged with
contaminants.
Net temperature or
Efficiency calculation
incorrect.
Ambient temperature set wrong during
Automatic Calibration.
Flue temperature readings
erratic
Temperature plug reversed in socket.
Faulty connection or break in cable or plug.
T flue or T nett displays
(- - - -)
Probe not connected.
X-Air, Efficiency or CO
2
display (- O>-)
Oxygen reading is above 18%