AUTOplus manual
Page 40
APPENDICES
A. MAIN DISPLAY PARAMETERS
The parameters and their meanings are detailed as follows: -
FUEL: The selected fuel will be displayed, i.e. PETROL.
PETROL Leaded or Unleaded petrol/gasoline.
LPG Liquid Petroleum Gas
CNG Compressed Natural Gas
DIESEL
DATE : Analyzer date. See Set-Up menu to change.
TIME : Analyzer time. Use Set-Up menu to change.
BATTERY : Displays the battery level from 0-100%. During normal data
display operation the analyzer will show LOW BATTERY briefly at
intervals of about 45 seconds for less than 30% of charge. The
analyzer will turn off if less than 10% charge is detected. With the
charger connected the display will read close to 100%.
Note: Allowing the battery to discharge fully may destroy it.
O2 : Oxygen measured in the exhaust gas indicated in percentage %.
With the pump off the analyzer will display - - - -. If there is a fault
with the oxygen sensor then FLT will be displayed.
CO : Carbon monoxide measured in the exhaust gas indicated in
percentage %. With the pump off the analyzer will display - - - -. If
there is a fault with the CO reading then FLT will be displayed.
CO2 : Carbon dioxide measured in the exhaust gas in percentage %.
With the pump off the analyzer will display - - - -. If there is a fault
with the CO
2
reading then FLT will be displayed.
HC : Hydrocarbons measured in the exhaust gas indicated in ppm
(parts per million) n-hexane (petrol). With the pump off the
analyzer will display - - - -. If there is a fault with the HC reading
then FLT will be displayed.