14
Electric Dispenser Hoist EWL 2, 3, 4
A 10 Operating faults
Power cord 9 does not coil up
automatically
Possible causes
a) Power cord 9 jammed
b) Cable pulley l defective
Remedy
a) Remove obstacle to the power
cord 9
b) Check cable pulley l and
replace if necessary
Steel 5 rope does not coil
Possible causes
a) Steel rope 5 twisted
b) Button 7 or motor faulty
Remedy
a) Unroll the steel rope 5 to the fullest
extent and roll it up again under load
b) Check button 7 or motor and
replace if necessary