English
6-1
Chapter
6
Specifications and
Standards
CAUTION
US Federal law restricts this device to sale by or on the order of a
dentist or other licensed practitioner.
Sensor Specifications
Sensor Dimensions KaVo IXS Size 1: 36.9 x 25.4 x 7.7 mm
KaVo IXS Size 2: 41.8 x 30.6 x 7.8 mm
Sensor Image Area Indirect converting dental intraoral X-ray sensor
1539 by 1026 pixels for Size 1
1842 by 1324 pixels for Size 2
19.5 μm pixel size
Resolution 20+ visible lp/mm
X-ray Parameters Sensor can be used with dental X-ray generators in
the range of 60 to 70 kV; at minimal 40 μGy incident
dose
Software Architecture Operating Systems Supported:
• Microsoft Windows 7 with current service pack
• Microsoft Windows 8.1 with current service pack
• Microsoft Windows 10 with current service pack
Electrical Rating DC 5V, 0.5 W
Connection to PC USB 2.0 compliant / USB 3.x compatible
Protection Against
Shock
Type BF
Applied Part
Mode of Operation Continuous
Method of Sterilization Sensor not suitable for sterilization
KaVo S1.book Page 1 Tuesday, August 6, 2019 2:36 PM